Anal Chim Acta ; 632(1): 93-100, 2009 Jan 19.
Artigo em Inglês | MEDLINE | ID: mdl-19100887

RESUMO

A highly selective sequential injection (SI) method for the automated determination of weak-acid-dissociable cyanides is reported. The analytical procedure is based on the on-line reaction of the analyte with ninhydrin in carbonate medium to form a coloured product (lambda(max)=510 nm). Cyanides are removed from sample matrix by acidification through a gas-diffusion step incorporated in the SI manifold. The effect of instrumental and chemical variables was studied. By adopting an on-line standard addition protocol, the sensitivity of the proposed method was enhanced drastically, without affecting the determination range. The assay was validated in terms of linearity (up to 200 microg L(-1)), limit of detection (c(L)=2.5 microg L(-1)), limit of quantitation (c(Q)=7.5 microg L(-1)), precision (s(r)<2.5% at 100 microg L(-1)) and selectivity. High tolerance against critical species such as sulfides and thiocyanates was achieved. The applicability of the method was demonstrated by analyzing tap and mineral water samples at levels below the limits established by international E.U. and U.S. organizations. The percent recoveries were satisfactory in all cases, ranging between 94.2 and 103.6%.