1.
Artigo em Inglês | MEDLINE | ID: mdl-23366920

RESUMO

A 64-channel Neuro-Stimulation-Recording chip named NRS64 for neural activity measurements has been designed and tested. The NRS64 occupies 5×5 mm² of silicon area and consumes only 25 µW/channel. A low cut-off frequency can be tuned in the 60 mHz-100 Hz range while a high cut-off frequency can be set to 4.7 kHz or 12 kHz. A voltage gain can be set to 139 V/V or 1100 V/V. A measured input referenced noise is 3.7 µV rms in 100 Hz-12 kHz band and 7.6 µV rms in 3 Hz-12 kHz band. A digital correction is used in each channel to tune the low cut-off frequency and offset voltage. Each channel is equipped additionally with a stimulation circuit with an artifact cancellation circuit. The stimulation circuit can be set with 8-bit resolution in six different ranges from 500 nA-512 µA range.


Assuntos
Potenciais de Ação/fisiologia , Amplificadores Eletrônicos , Conversão Análogo-Digital , Estimulação Elétrica/instrumentação , Eletroencefalografia/instrumentação , Processamento de Sinais Assistido por Computador/instrumentação , Fontes de Energia Elétrica , Desenho de Equipamento , Análise de Falha de Equipamento , Semicondutores , Razão Sinal-Ruído