Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros










Base de dados
Intervalo de ano de publicação
1.
J Diabetes Sci Technol ; 17(1): 125-132, 2023 01.
Artigo em Inglês | MEDLINE | ID: mdl-34515537

RESUMO

BACKGROUND: Annual peaks in temperature and humidity exceed manufacturers' specifications for blood glucose test strip storage in Thailand. Health Promoting Hospitals (HPH) do not always provide the same level of health services that hospitals do since they often only turn on air-conditioning units during working hours. METHODS: The blood glucose testing performance of 4 glucose meters were investigated for short and long terms stress at 5 simulated conditions. Temperature and relative humidity (RH) at 5 HPHs in different regions of Thailand were monitored for 9 weeks during April to July 2019. The use of air conditioning in storage rooms for storing test strips was surveyed at 131 HPHs using questionnaires. RESULTS: Median-paired differences of glucose measurements with stressed test strips in 5 simulated conditions significantly differed (P < 0.05) both in the short term (3 days) and in the long term (30 days) with 4 glucose meters when compared to unstressed test strips. The average temperature of all HPHs exceeded 30°C (86°F). The average RH was 84%. There was only one HPH that occasionally turned on its air conditioners. Most HPHs kept both opened and unopened vials of test strips in rooms without air conditioners. Further, 21.4-32.0% of HPHs kept test strips at room with air conditioners. CONCLUSIONS: This study provides evidence for poor performance of blood glucose testing by glucose meters that are affected by adverse environmental conditions. The environmental for test strips storage at HPHs should be considered to prevent analytical errors of glucose measurement.


Assuntos
Glicemia , Glucose , Humanos , Fitas Reagentes , Hospitais , Promoção da Saúde
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...