Radiat Prot Dosimetry ; 164(3): 361-7, 2015 Apr.
Artigo em Inglês | MEDLINE | ID: mdl-25213263

RESUMO

This study aimed to validate a MOSFET dosemeter system for determining absorbed and effective doses (EDs) in the dose and energy range used in diagnostic radiology. Energy dependence, dose linearity and repeatability of the dosemeter were examined. The absorbed doses (ADs) were compared at anterior-posterior projection and the EDs were determined at posterior-anterior, anterior-posterior and lateral projections of thoracic imaging using an anthropomorphic phantom. The radiation exposures were made using digital radiography systems. This study revealed that the MOSFET system with high sensitivity bias supply set-up is sufficiently accurate for AD and ED determination. The dosemeter is recommended to be calibrated for energies <60 and >80 kVp. The entrance skin dose level should be at least 5 mGy to minimise the deviation of the individual dosemeter dose. For ED determination, dosemeters should be implanted perpendicular to the surface of the phantom to prevent the angular dependence error.


Assuntos
Imagens de Fantasmas , Monitoramento de Radiação/instrumentação , Intensificação de Imagem Radiográfica/instrumentação , Radiologia/instrumentação , Pele/efeitos da radiação , Calibragem , Humanos , Método de Monte Carlo , Equipamentos de Proteção , Doses de Radiação