Radiat Prot Dosimetry ; 198(12): 815-820, 2022 Aug 26.
Artigo em Inglês | MEDLINE | ID: mdl-35718757

RESUMO

The main purpose of this study was to determine local diagnostic reference level (LDRL) for routine computed tomography (CT) examination in Bafoussam, western Cameroon. The exposure parameters and dose quantities were collected manually. This retrospective, evaluative and comparative study was conducted to determine LDRLs for routine head CT examination in Bafoussam, to optimize these procedures in the region. The 75th percentile values of the calculated volume CT dose index (CTDIvol) and dose length product (DLP) were proposed as LDRL. The sample was classified in four age groups: < 1, 1-5, 5-10 and 10-15 y. The LDRLs obtained for the four age groups were: 24 mGy and 381.32 mGy.cm, 42.5 mGy and 875.55 mGy.cm, 45.85 mGy and 939.62 mGy.cm, 57.12 mGy and 1222.3 mGy.cm, respectively. The 75th percentile CTDIvol and DLP dose values for this study are higher than international values. We propose a coordinating discussions and collaboration about patient's and specific equipment's change information's, between radiologists, medical imaging technicians and medical physicist, which can reduce absorbed doses and improved medical practice in hospitals.


Assuntos
Níveis de Referência de Diagnóstico , Tomografia Computadorizada por Raios X , Camarões , Criança , Humanos , Doses de Radiação , Valores de Referência , Estudos Retrospectivos , Tomografia Computadorizada por Raios X/métodos