RESUMO
Nanoscale magnetic resonance imaging (NanoMRI) is an active area of applied research with potential applications in structural biology and quantum engineering. The success of this technological vision hinges on improving the instrument's sensitivity and functionality. A particular challenge is the optimization of the magnetic field gradient required for spatial encoding and of the radio frequency field used for spin control, in analogy to the components used in clinical MRI. In this work, we present the fabrication and characterization of a magnet-in-microstrip device that yields a compact form factor for both elements. We find that our design leads to a number of advantages, among them a 4-fold increase of the magnetic field gradient compared to those achieved with traditional fabrication methods. Our results can be useful for boosting the efficiency of a variety of different experimental arrangements and detection principles in the field of NanoMRI.
RESUMO
Hexadecane exhibits pronounced molecular layering upon confinement to gaps of a few nanometer width which is discussed for its role in boundary lubrication. We have probed the mechanical properties of the confined layers with the help of an atomic force microscope, by quasi-static normal force measurements and by analyzing the lateral tip motion of a magnetically actuated torsional cantilever oscillation. The molecular layering is modeled by a oscillatory force curve and the tip approach is simulated assuming thermal equilibrium correlations in the liquid. The shear response of the confined layers reveals gradually increasing stiffness and viscous dissipation for a decreasing number of confined layers.