Rev Sci Instrum ; 91(6): 065109, 2020 Jun 01.
Artigo em Inglês | MEDLINE | ID: mdl-32611059

RESUMO

We present a multi-purpose radiation furnace designed for x-ray experiments at synchrotrons. The furnace is optimized specifically for dark-field x-ray microscopy (DFXM) of crystalline materials at beamline ID06 of the European Synchrotron Radiation Facility. The furnace can reach temperatures above 1200 °C with a thermal stability better than 10 °C, with heating and cooling rates up to 30 K/s. The non-contact heating design enables samples to be heated either in air or in a controlled atmosphere contained within a capillary tube. The temperature was calibrated via the thermal expansion of an α-iron grain. Temperature profiles in the y and z axes were measured by scanning a thermocouple through the focal spot of the radiation furnace. In the current configuration of the beamline, this furnace can be used for DFXM, near-field x-ray topography, bright-field x-ray nanotomography, high-resolution reciprocal space mapping, and limited powder diffraction experiments. As a first application, we present a DFXM case study on isothermal heating of a commercially pure single crystal of aluminum.