Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros










Base de dados
Intervalo de ano de publicação
1.
Radiat Prot Dosimetry ; 111(3): 297-304, 2004.
Artigo em Inglês | MEDLINE | ID: mdl-15266080

RESUMO

The purpose of the study was to evaluate the dose performance of a flat-panel (FP) and an image intensifier (II) charge coupled device (CCD) digital fluoroscopy X-ray systems newly installed in an Interventional Cardiology (IC) department. Filter entrance dose rate, detector dose rate (during fluoroscopy) and filter entrance dose per image were measured at 70 cm from the focus using 2 mm copper sheets to mimic normal size patient. Image quality was also evaluated. The patient dose survey included 277 patients, which had either a Coronary Angiography (CA) or a Percutaneous Transluminal Coronary Angioplasty (PTCA). Dose area product (DAP), fluoroscopy time (T) and total number of frames (F) values were also collected. The results showed that both systems performed within international recommendations with the exception of higher cine radiation doses, stressing the fact that neither specific protocols of measurement nor reference values for digital equipment were provided by the official bodies.


Assuntos
Cardiologia/instrumentação , Análise de Falha de Equipamento/métodos , Segurança de Equipamentos/métodos , Fluoroscopia/instrumentação , Intensificação de Imagem Radiográfica/instrumentação , Radiografia Intervencionista/instrumentação , Radiometria/métodos , Ecrans Intensificadores para Raios X , Doses de Radiação , Proteção Radiológica/métodos , Medição de Risco/métodos , Fatores de Risco , Raios X
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...