Stud Health Technol Inform ; 189: 101-6, 2013.
Artigo em Inglês | MEDLINE | ID: mdl-23739366

RESUMO

The paper presents a non-invasive method and system for a long-term and continuous monitoring of the central aortic pressure (CAP) waveform and the augmentation index (AI). The CAP curve is estimated from the measured radial electrical bio-impedance (EBI) using spectral domain transfer functions (TF), which are established on the basis of data analysis during clinical experiments. Experiments were carried out on 3 volunteers by now. During the experiment, a 0.5 mg sublingual nitroglycerin tablet was administrated to each volunteer. Both, the reconstructed CAP curve and the AI have very good correlation with the results obtained by the SphygmoCor system. But, in opposite to the traditional tonometry based CAP curve and AI estimation methods, the proposed one is more convenient to use and allows continuous and long-term personalized monitoring of the CAP curve and of the AI.


Assuntos
Pressão Arterial , Determinação da Pressão Arterial/instrumentação , Diagnóstico por Computador/instrumentação , Hipertensão/diagnóstico , Hipertensão/fisiopatologia , Monitorização Ambulatorial/instrumentação , Pletismografia de Impedância/instrumentação , Desenho de Equipamento , Análise de Falha de Equipamento , Feminino , Humanos , Masculino , Reprodutibilidade dos Testes , Sensibilidade e Especificidade , Adulto Jovem