Am J Cardiol ; 96(3): 414-6, 2005 Aug 01.
Artigo em Inglês | MEDLINE | ID: mdl-16054471

RESUMO

We performed a retrospective analysis of 250 records of consecutive, newly implanted, pacemaker patients from a single center to determine the rate of postimplant complications and observations discovered before and during the prehospital discharge evaluation. No observations occurred in 246 of 250 patients (98.4%) (1-sided 95% confidence interval 96.4%). Of the 250 patients, 4 had observations that were discovered at the prehospital discharge check and required reprogramming to increase the sensitivity safety margin (3 atrial and 1 ventricular). We documented only 1 complication that was discovered before the predischarge evaluation through telemetry and resulted in an atrial lead revision.