Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros










Base de dados
Intervalo de ano de publicação
1.
Rev Sci Instrum ; 87(7): 073706, 2016 Jul.
Artigo em Inglês | MEDLINE | ID: mdl-27475564

RESUMO

A multi-wavelength, high contrast contact radiography system has been developed to characterize density variations in ultra-low density aerogel foams. These foams are used to generate a ramped pressure drive in materials strength experiments at the National Ignition Facility and require precision characterization in order to reduce errors in measurements. The system was used to characterize density variations in carbon and silicon based aerogels to ∼10.3% accuracy with ∼30 µm spatial resolution. The system description, performance, and measurement results collected using a 17.8 mg/cc carbon based JX-6 (C20H30) aerogel are discussed in this manuscript.

SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...