1.
JBJS Case Connect ; 11(2)2021 06 24.
Artigo em Inglês | MEDLINE | ID: mdl-34166254

RESUMO

CASE: We describe 2 cases of gross trunnion failure (GTF) in males with Centers for Disease Control and Prevention Class 3 obesity 10 years after metal-on-polyethylene total hip arthroplasty (THA) with a titanium-alloy femoral stem and Type 1 taper. One patient received a large diameter cobalt-chromium femoral head, whereas the other received a smaller diameter head, both with high-offset femoral stems. CONCLUSION: This is the first report of GTF involving the Echo Bi-Metric femoral stem after metal-on-polyethylene THA, and surgeons should consider the potential complication of GTF when using this specific femoral stem with metal heads in obese male patients.


Assuntos
Artroplastia de Quadril , Prótese de Quadril , Artroplastia de Quadril/efeitos adversos , Prótese de Quadril/efeitos adversos , Humanos , Masculino , Polietileno , Desenho de Prótese , Falha de Prótese , Estados Unidos