1.
Artigo em Inglês | MEDLINE | ID: mdl-31745534

RESUMO

The Kretzfile format is used to store 3D ultrasound data, from GE Voluson ultrasound scanners. The geometry used in these hies is a toroidal coordinate system. Cartesian coordinates are required to allow application of advanced image libraries like ITK and scikit-image. We present ITK transformation and utilities to convert Kretzfiles to cartesian coordinates. Previous work (SlicerHeart, 2017) has enabled the reading of kretz files and approximate coordinate transformations. This work will enable medical imaging researchers to investigate clinically 3D ultrasound.

2.
Phys Med Biol ; 62(3): 858-877, 2017 02 07.
Artigo em Inglês | MEDLINE | ID: mdl-28072582

RESUMO

Digital breast tomosynthesis (DBT) is under consideration to replace or to be used in combination with 2D-mammography in breast screening. The aim of this study was the comparison of the detection of microcalcification clusters by human observers in simulated breast images using 2D-mammography, narrow angle (15°/15 projections) and wide angle (50°/25 projections) DBT. The effects of the cluster height in the breast and the dose to the breast on calcification detection were also tested. Simulated images of 6 cm thick compressed breasts were produced with and without microcalcification clusters inserted, using a set of image modelling tools for 2D-mammography and DBT. Image processing and reconstruction were performed using commercial software. A series of 4-alternative forced choice (4AFC) experiments was conducted for signal detection with the microcalcification clusters as targets. Threshold detectable calcification diameter was found for each imaging modality with standard dose: 2D-mammography: 2D-mammography (165 ± 9 µm), narrow angle DBT (211 ± 11 µm) and wide angle DBT (257 ± 14 µm). Statistically significant differences were found when using different doses, but different geometries had a greater effect. No differences were found between the threshold detectable calcification diameters at different heights in the breast. Calcification clusters may have a lower detectability using DBT than 2D imaging.


Assuntos
Doenças Mamárias/diagnóstico por imagem , Calcinose/diagnóstico por imagem , Mamografia/métodos , Imagens de Fantasmas , Intensificação de Imagem Radiográfica/métodos , Interpretação de Imagem Radiográfica Assistida por Computador/métodos , Tomografia por Raios X/métodos , Feminino , Humanos , Intensificação de Imagem Radiográfica/instrumentação , Software
3.
Phys Med ; 32(4): 568-74, 2016 Apr.
Artigo em Inglês | MEDLINE | ID: mdl-27061872

RESUMO

PURPOSE: To investigate the relationship between image quality measurements and the clinical performance of digital mammographic systems. METHODS: Mammograms containing subtle malignant non-calcification lesions and simulated malignant calcification clusters were adapted to appear as if acquired by four types of detector. Observers searched for suspicious lesions and gave these a malignancy score. Analysis was undertaken using jackknife alternative free-response receiver operating characteristics weighted figure of merit (FoM). Images of a CDMAM contrast-detail phantom were adapted to appear as if acquired using the same four detectors as the clinical images. The resultant threshold gold thicknesses were compared to the FoMs using a linear regression model and an F-test was used to find if the gradient of the relationship was significantly non-zero. RESULTS: The detectors with the best image quality measurement also had the highest FoM values. The gradient of the inverse relationship between FoMs and threshold gold thickness for the 0.25mm diameter disk was significantly different from zero for calcification clusters (p=0.027), but not for non-calcification lesions (p=0.11). Systems performing just above the minimum image quality level set in the European Guidelines for Quality Assurance in Breast Cancer Screening and Diagnosis resulted in reduced cancer detection rates compared to systems performing at the achievable level. CONCLUSIONS: The clinical effectiveness of mammography for the task of detecting calcification clusters was found to be linked to image quality assessment using the CDMAM phantom. The European Guidelines should be reviewed as the current minimum image quality standards may be too low.


Assuntos
Neoplasias da Mama/diagnóstico por imagem , Mamografia/métodos , Neoplasias da Mama/metabolismo , Neoplasias da Mama/patologia , Calcinose/diagnóstico por imagem , Calcinose/metabolismo , Calcinose/patologia , Feminino , Guias como Assunto , Humanos , Mamografia/normas , Intensificação de Imagem Radiográfica/métodos
4.
Eur Radiol ; 26(3): 874-83, 2016 Mar.
Artigo em Inglês | MEDLINE | ID: mdl-26105023

RESUMO

OBJECTIVE: To compare the performance of different types of detectors in breast cancer detection. METHODS: A mammography image set containing subtle malignant non-calcification lesions, biopsy-proven benign lesions, simulated malignant calcification clusters and normals was acquired using amorphous-selenium (a-Se) detectors. The images were adapted to simulate four types of detectors at the same radiation dose: digital radiography (DR) detectors with a-Se and caesium iodide (CsI) convertors, and computed radiography (CR) detectors with a powder phosphor (PIP) and a needle phosphor (NIP). Seven observers marked suspicious and benign lesions. Analysis was undertaken using jackknife alternative free-response receiver operating characteristics weighted figure of merit (FoM). The cancer detection fraction (CDF) was estimated for a representative image set from screening. RESULTS: No significant differences in the FoMs between the DR detectors were measured. For calcification clusters and non-calcification lesions, both CR detectors' FoMs were significantly lower than for DR detectors. The calcification cluster's FoM for CR NIP was significantly better than for CR PIP. The estimated CDFs with CR PIP and CR NIP detectors were up to 15% and 22% lower, respectively, than for DR detectors. CONCLUSION: Cancer detection is affected by detector type, and the use of CR in mammography should be reconsidered. KEY POINTS: The type of mammography detector can affect the cancer detection rates. CR detectors performed worse than DR detectors in mammography. Needle phosphor CR performed better than powder phosphor CR. Calcification clusters detection is more sensitive to detector type than other cancers.


Assuntos
Neoplasias da Mama/diagnóstico por imagem , Calcinose/diagnóstico por imagem , Mamografia/instrumentação , Idoso , Detecção Precoce de Câncer/instrumentação , Detecção Precoce de Câncer/métodos , Feminino , Humanos , Mamografia/métodos , Programas de Rastreamento/instrumentação , Programas de Rastreamento/métodos , Pessoa de Meia-Idade , Agulhas , Variações Dependentes do Observador , Curva ROC , Intensificação de Imagem Radiográfica/métodos