1.
J Neural Eng ; 9(4): 046014, 2012 Aug.
Artigo em Inglês | MEDLINE | ID: mdl-22791690

RESUMO

The objective of this work is to develop and test a photovoltaic retinal prosthesis for restoring sight to patients blinded by degenerative retinal diseases. A silicon photodiode array for subretinal stimulation has been fabricated by a silicon-integrated-circuit/MEMS process. Each pixel in the two-dimensional array contains three series-connected photodiodes, which photovoltaically convert pulsed near-infrared light into bi-phasic current to stimulate nearby retinal neurons without wired power connections. The device thickness is chosen to be 30 µm to absorb a significant portion of light while still being thin enough for subretinal implantation. Active and return electrodes confine current near each pixel and are sputter coated with iridium oxide to enhance charge injection levels and provide a stable neural interface. Pixels are separated by 5 µm wide trenches to electrically isolate them and to allow nutrient diffusion through the device. Three sizes of pixels (280, 140 and 70 µm) with active electrodes of 80, 40 and 20 µm diameter were fabricated. The turn-on voltages of the one-diode, two-series-connected diode and three-series-connected diode structures are approximately 0.6, 1.2 and 1.8 V, respectively. The measured photo-responsivity per diode at 880 nm wavelength is ∼0.36 A W(-1), at zero voltage bias and scales with the exposed silicon area. For all three pixel sizes, the reverse-bias dark current is sufficiently low (<100 pA) for our application. Pixels of all three sizes reliably elicit retinal responses at safe near-infrared light irradiances, with good acceptance of the photodiode array in the subretinal space. The fabricated device delivers efficient retinal stimulation at safe near-infrared light irradiances without any wired power connections, which greatly simplifies the implantation procedure. Presence of the return electrodes in each pixel helps to localize the current, and thereby improves resolution.


Assuntos
Implantação de Prótese/instrumentação , Implantação de Prótese/métodos , Próteses Visuais , Animais , Microeletrodos , Estimulação Luminosa/métodos , Ratos , Ratos Long-Evans , Silício/administração & dosagem , Suínos
2.
IEEE Trans Biomed Circuits Syst ; 5(5): 468-80, 2011 Oct.
Artigo em Inglês | MEDLINE | ID: mdl-23852178

RESUMO

Photodiode circuits show promise for the development of high-resolution retinal prostheses. While several of these systems have been constructed and some even implanted in humans, existing descriptions of the complex optoelectronic interaction between light, photodiode, and the electrode/electrolyte load are limited. This study examines this interaction in depth with theoretical calculations and experimental measurements. Actively biased photoconductive and passive photovoltaic circuits are investigated, with the photovoltaic circuits consisting of one or more diodes connected in series, and the photoconductive circuits consisting of a single diode in series with a pulsed bias voltage. Circuit behavior and charge injection levels were markedly different for platinum and sputtered iridium-oxide film (SIROF) electrodes. Photovoltaic circuits were able to deliver 0.038 mC/cm(2) (0.75 nC/phase) per photodiode with 50- µm platinum electrodes, and 0.54-mC/cm(2) (11 nC/phase) per photodiode with 50-µ m SIROF electrodes driven with 0.5-ms pulses of light at 25 Hz. The same pulses applied to photoconductive circuits with the same electrodes were able to deliver charge injections as high as 0.38 and 7.6 mC/cm(2) (7.5 and 150 nC/phase), respectively. We demonstrate photovoltaic stimulation of rabbit retina in-vitro, with 0.5-ms pulses of 905-nm light using peak irradiance of 1 mW/mm(2). Based on the experimental data, we derive electrochemical and optical safety limits for pixel density and charge injection in various circuits. While photoconductive circuits offer smaller pixels, photovoltaic systems do not require an external bias voltage. Both classes of circuits show promise for the development of high-resolution optoelectronic retinal prostheses.

3.
J Neural Eng ; 4(1): S72-84, 2007 Mar.
Artigo em Inglês | MEDLINE | ID: mdl-17325419

RESUMO

The design of high-resolution retinal prostheses presents many unique engineering and biological challenges. Ever smaller electrodes must inject enough charge to stimulate nerve cells, within electrochemically safe voltage limits. Stimulation sites should be placed within an electrode diameter from the target cells to prevent 'blurring' and minimize current. Signals must be delivered wirelessly from an external source to a large number of electrodes, and visual information should, ideally, maintain its natural link to eye movements. Finally, a good system must have a wide range of stimulation currents, external control of image processing and the option of either anodic-first or cathodic-first pulses. This paper discusses these challenges and presents solutions to them for a system based on a photodiode array implant. Video frames are processed and imaged onto the retinal implant by a head-mounted near-to-eye projection system operating at near-infrared wavelengths. Photodiodes convert light into pulsed electric current, with charge injection maximized by applying a common biphasic bias waveform. The resulting prosthesis will provide stimulation with a frame rate of up to 50 Hz in a central 10 degrees visual field, with a full 30 degrees field accessible via eye movements. Pixel sizes are scalable from 100 to 25 microm, corresponding to 640-10,000 pixels on an implant 3 mm in diameter.


Assuntos
Terapia por Estimulação Elétrica/instrumentação , Eletrodos Implantados , Eletrônica Médica/instrumentação , Óptica e Fotônica/instrumentação , Próteses e Implantes , Doenças Retinianas/reabilitação , Gravação em Vídeo/instrumentação , Animais , Inteligência Artificial , Terapia por Estimulação Elétrica/métodos , Análise de Falha de Equipamento , Humanos , Interpretação de Imagem Assistida por Computador/instrumentação , Microeletrodos , Desenho de Prótese , Retina/cirurgia , Terapia Assistida por Computador/instrumentação , Terapia Assistida por Computador/métodos , Gravação em Vídeo/métodos