1.
AAPS PharmSci ; 3(3): E23, 2001.
Artigo em Inglês | MEDLINE | ID: mdl-11741274

RESUMO

The objective of this study was to demonstrate the use of transmission Fourier transform near-infrared (FT-NIR) spectroscopy for quantitative analysis of an active ingredient in a translucent gel formulation. Gels were prepared using Carbopol 980 with 0%, 1%, 2%, 4%, 6%, and 8% ketoprofen and analyzed with an FT-NIR spectrophotometer operated in the transmission mode. The correlation coefficient of the calibration was 0.9996, and the root mean squared error of calibration was 0.0775%. The percent relative standard deviation for multiple measurements was 0.10%. The results prove that FT-NIR can be a good alternative to other, more time-consuming means of analysis for these types of formulations.


Assuntos
Anti-Inflamatórios não Esteroides/análise , Cetoprofeno/análise , Administração Tópica , Anti-Inflamatórios não Esteroides/administração & dosagem , Análise de Fourier , Géis , Cetoprofeno/administração & dosagem , Espectroscopia de Luz Próxima ao Infravermelho