Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros










Base de dados
Intervalo de ano de publicação
1.
Eur J Clin Microbiol Infect Dis ; 40(4): 699-705, 2021 Apr.
Artigo em Inglês | MEDLINE | ID: mdl-33034779

RESUMO

Delayed entry of patient blood culture samples into a microbial detection system is unavoidable at times, due to off-shift staffing or transporting samples to centralized laboratories. Pre-incubation time and temperature of blood culture bottles are the most critical factors impacting recovery and detection of microorganisms. A total of 1377 BACT/ALERT® (BTA) Fastidious Antimicrobial Neutralization (FAN® PLUS) bottles (FA PLUS, FN PLUS, and PF PLUS) were tested after delayed entry times of 24 and 36 h at 20-25 °C (room temperature, RT) prior to loading into the BACT/ALERT® VIRTUO® microbial detection system (VIRTUO). Clinically relevant organisms were inoculated into bottles with 5-84 colony forming units (CFU) per bottle, and human blood (0 to 10 mL), and then loaded into the VIRTUO. When bottles were loaded without delay, a mean time to detection (TTD) of 9.6 h was observed. For delayed bottles, the TTD reported by the VIRTUO was added to the 24-h and 36-h delay times and resulted in average time to results of 32.5 h and 42.5 h, respectively. The FAN PLUS bottles in conjunction with the VIRTUO produced acceptable results when delays up to 24 h at 20-25 °C occur in loading.


Assuntos
Técnicas Bacteriológicas/métodos , Hemocultura/métodos , Manejo de Espécimes , Meios de Cultura , Humanos , Fatores de Tempo
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...