Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros










Base de dados
Intervalo de ano de publicação
1.
Talanta ; 81(3): 1081-7, 2010 May 15.
Artigo em Inglês | MEDLINE | ID: mdl-20298897

RESUMO

A new method based on corona discharge ion mobility spectrometry (CD-IMS) was developed for the analysis of biogenic amines including spermidine, spermine, putrescine, and cadaverine. The ion mobility spectra of the compounds were obtained with and without n-Nonylamine used as the reagent gas. The high proton affinity of n-Nonylamine prevented ion formation from compounds with a proton affinity lower than that of n-Nonylamine and, therefore, enhanced its selectivity. It was also realized that the ion mobility spectrum of n-Nonylamine varied with its concentration. A sample injection port of a gas chromatograph was modified and used as the sample introduction system into the CD-IMS. The detection limits, dynamic ranges, and analytical parameters of the compounds with and without using the reagent gas were obtained. The detection limits and dynamic ranges of the compounds were about 2ng and 2 orders of magnitude, respectively. The wide dynamic range of CD-IMS originates from the high current of the corona discharge. The results revealed the high capability of the CD-IMS for the analysis of biogenic amines.


Assuntos
Aminas Biogênicas/análise , Técnicas de Química Analítica , Espectrometria de Massas/instrumentação , Análise Espectral/métodos , Calibragem , Gases , Humanos , Íons , Espectrometria de Massas/métodos , Metanol/química , Modelos Químicos , Prótons , Reprodutibilidade dos Testes , Água/química
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...