1.
Appl Radiat Isot ; 52(5): 1203-7, 2000 May.
Artigo em Inglês | MEDLINE | ID: mdl-10836431

RESUMO

Alanine/EPR dosimetry was applied to measure dose distributions during electron arc therapy treatments generated by electron beams moving isocentrically over a stated arc. Alanine-polyethylene pellets, prepared at ISS according to the NIST recipe, inserted in a homogeneous cylindrical and anthropomorphic phantoms, were used. Preliminary, alanine response to static electron beams in the (6-20) MeV nominal energy range was studied. Then alanine dosimetry was applied to determine the dose versus the gantry angle at a reference point in a cylindrical homogeneous phantom. Finally, arc therapy treatment planning verification was performed in anthropomorphic phantom.


Assuntos
Espectroscopia de Ressonância de Spin Eletrônica/métodos , Elétrons/uso terapêutico , Planejamento da Radioterapia Assistida por Computador/métodos , Alanina/efeitos da radiação , Humanos , Imagens de Fantasmas , Polietileno , Radioterapia de Alta Energia