Med Phys ; 40(11): 111706, 2013 Nov.
Artigo em Inglês | MEDLINE | ID: mdl-24320414

RESUMO

PURPOSE: Accurate treatment delivery in high dose rate (HDR) brachytherapy requires correct source dwell positions and dwell times to be administered relative to each other and to the surrounding anatomy. Treatment delivery inaccuracies predominantly occur for two reasons: (i) anatomical movement or (ii) as a result of human errors that are usually related to incorrect implementation of the planned treatment. Electronic portal imaging devices (EPIDs) were originally developed for patient position verification in external beam radiotherapy and their application has been extended to provide dosimetric information. The authors have characterized the response of an EPID for use with an (192)Ir brachytherapy source to demonstrate its use as a verification device, providing both source position and dosimetric information. METHODS: Characterization of the EPID response using an (192)Ir brachytherapy source included investigations of reproducibility, linearity with dose rate, photon energy dependence, and charge build-up effects associated with exposure time and image acquisition time. Source position resolution in three dimensions was determined. To illustrate treatment verification, a simple treatment plan was delivered to a phantom and the measured EPID dose distribution compared with the planned dose. RESULTS: The mean absolute source position error in the plane parallel to the EPID, for dwells measured at 50, 100, and 150 mm source to detector distances (SDD), was determined to be 0.26 mm. The resolution of the z coordinate (perpendicular distance from detector plane) is SDD dependent with 95% confidence intervals of ± 0.1, ± 0.5, and ± 2.0 mm at SDDs of 50, 100, and 150 mm, respectively. The response of the EPID is highly linear to dose rate. The EPID exhibits an over-response to low energy incident photons and this nonlinearity is incorporated into the dose calibration procedure. A distance (spectral) dependent dose rate calibration procedure has been developed. The difference between measured and planned dose is less than 2% for 98.0% of pixels in a two-dimensional plane at an SDD of 100 mm. CONCLUSIONS: Our application of EPID dosimetry to HDR brachytherapy provides a quality assurance measure of the geometrical distribution of the delivered dose as well as the source positions, which is not possible with any current HDR brachytherapy verification system.


Assuntos
Braquiterapia/métodos , Radiometria/métodos , Algoritmos , Calibragem , Catéteres , Desenho de Equipamento , Humanos , Radioisótopos de Irídio/química , Erros Médicos/prevenção & controle , Movimento , Controle de Qualidade , Dosagem Radioterapêutica , Planejamento da Radioterapia Assistida por Computador/métodos , Reprodutibilidade dos Testes
2.
Australas Phys Eng Sci Med ; 35(2): 159-64, 2012 Jun.
Artigo em Inglês | MEDLINE | ID: mdl-22581501

RESUMO

Variations in anatomical delineation, principally due to a combination of inter-observer contributions and image-specificity, remain one of the most significant impediments to geometrically-accurate radiotherapy. Quantification of spatial variability of the delineated contours comprising a structure can be made with a variety of metrics, and the availability of software tools to apply such metrics to data collected during inter-observer or repeat-imaging studies would allow their validation. A suite of such tools have been developed which use an Extensible Markup Language format for the exchange of delineated 3D structures with radiotherapy planning or review systems. These tools provide basic operations for manipulating and operating on individual structures and related structure sets, and for deriving statistics on spatial variations of contours that can be mapped onto the surface of a reference structure. Use of these tools on a sample dataset is demonstrated together with import and display of results in the SWAN treatment plan review system.


Assuntos
Algoritmos , Reconhecimento Automatizado de Padrão/métodos , Neoplasias da Próstata/diagnóstico por imagem , Intensificação de Imagem Radiográfica/métodos , Interpretação de Imagem Radiográfica Assistida por Computador/métodos , Software , Humanos , Masculino , Reprodutibilidade dos Testes , Sensibilidade e Especificidade
3.
Australas Phys Eng Sci Med ; 35(1): 93-100, 2012 Mar.
Artigo em Inglês | MEDLINE | ID: mdl-22119931

RESUMO

A commercially available motion phantom (QUASAR, Modus Medical) was modified for programmable motion control with the aim of reproducing patient respiratory motion in one dimension in both the anterior-posterior and superior-inferior directions, as well as, providing controllable breath-hold and sinusoidal patterns for the testing of radiotherapy gating systems. In order to simulate realistic patient motion, the DC motor was replaced by a stepper motor. A separate 'chest-wall' motion platform was also designed to accommodate a variety of surrogate marker systems. The platform employs a second stepper motor that allows for the decoupling of the chest-wall and insert motion. The platform's accuracy was tested by replicating patient traces recorded with the Varian real-time position management (RPM) system and comparing the motion platform's recorded motion trace with the original patient data. Six lung cancer patient traces recorded with the RPM system were uploaded to the motion platform's in-house control software and subsequently replicated through the phantom motion platform. The phantom's motion profile was recorded with the RPM system and compared to the original patient data. Sinusoidal and breath-hold patterns were simulated with the motion platform and recorded with the RPM system to verify the systems potential for routine quality assurance of commercial radiotherapy gating systems. There was good correlation between replicated and actual patient data (P 0.003). Mean differences between the location of maxima in replicated and patient data-sets for six patients amounted to 0.034 cm with the corresponding minima mean equal to 0.010 cm. The upgraded motion phantom was found to replicate patient motion accurately as well as provide useful test patterns to aid in the quality assurance of motion management methods and technologies.


Assuntos
Neoplasias/radioterapia , Imagens de Fantasmas , Técnicas de Imagem de Sincronização Respiratória/métodos , Algoritmos , Simulação por Computador , Humanos , Movimento (Física) , Planejamento da Radioterapia Assistida por Computador , Software
4.
Med Phys ; 37(6): 2638-44, 2010 Jun.
Artigo em Inglês | MEDLINE | ID: mdl-20632575

RESUMO

The potential for detrimental incidents and the ever increasing complexity of patient treatments emphasize the need for accurate dosimetric verification in radiotherapy. For this reason, all curative treatments are verified, either pretreatment or in vivo, by electronic portal imaging device (EPID) dosimetry in the Radiation Oncology Department of The Netherlands Cancer Institute-Antoni van Leeuwenhoek hospital, Amsterdam, The Netherlands. Since the clinical introduction of the method in January 2005 until August 2009, treatment plans of 4337 patients have been verified. Among these plans, 17 serious errors were detected that led to intervention. Due to their origin, nine of these errors would not have been detected with pretreatment verification. The method is illustrated in detail by the case of a plan transfer error detected in a 5 x 5 Gy intensity-modulated radiotherapy (IMRT) rectum treatment. The EPID reconstructed dose at the isocenter was 6.3% below the planned value. Investigation of the plan transfer chain revealed that due to a network transfer error, the plan was corrupted. 3D analysis of the acquired EPID data revealed serious underdosage of the planning target volume: On average 11.6%, locally up to 20%. This report shows the importance of in vivo (EPID) dosimetry for all treatment plans as well as the ability of the method to assess the dosimetric impact of deviations found.


Assuntos
Algoritmos , Radiometria/métodos , Planejamento da Radioterapia Assistida por Computador/métodos , Radioterapia Conformacional/métodos , Tomografia Computadorizada por Raios X/métodos , Ecrans Intensificadores para Raios X , Humanos , Dosagem Radioterapêutica , Reprodutibilidade dos Testes , Sensibilidade e Especificidade
5.
Phys Med Biol ; 55(3): 873-81, 2010 Feb 07.
Artigo em Inglês | MEDLINE | ID: mdl-20071767

RESUMO

Despite the highly localized doses that may be delivered via stereotactic radiotherapy, a small dose is nonetheless delivered to out-of-field regions, which may cause detriment to the patient. In this work, a systematic set of dose measurements have been undertaken up to a distance of 45 cm from the isocentre, for stereotactic fields shaped by a BrainLAB mini-multileaf collimator (MMLC) mounted on a Varian 600C linear accelerator. A range of treatment parameters were varied so as to determine the factors of greatest influence and establish relationships with dose. The commercial treatment planning software (TPS) miscalculates the dose to out-of-field regions. Measured dose decreases consistently out to 45 cm, whereas the TPS decreases out to 10-15 cm, at which point the predicted dose is constant. At the 5-10 cm off-axis distance (OAD), measurements indicate doses of about 5-10% of the dose at the isocentre, 1% at 15 cm OAD and 0.1% at 45 cm OAD. There are several observed trends. Greater MMLC field sizes (with static jaw) result in higher out-of-field dose, as do shallower depths. The source-to-surface distance does not greatly influence peripheral dose. However, the results given in this work do indicate that simple treatment arrangements, such as preferable collimator rotation, would in certain cases reduce out-of-field dose by an order of magnitude. Peripheral dose raises questions of treatment optimization, particularly in cases where patients have a long life expectancy in which secondary effects may become manifest, such as in the treatment of paediatric patients or those with a non-malignant primary. For instance, for a 20 Gy hypo-fractionated treatment, dose to out-of-field regions is of the order of cGy-a substantial dose in radiation protection terms.


Assuntos
Radiometria/métodos , Radiocirurgia/instrumentação , Dosagem Radioterapêutica , Planejamento da Radioterapia Assistida por Computador/métodos , Humanos , Aceleradores de Partículas/instrumentação , Imagens de Fantasmas , Planejamento da Radioterapia Assistida por Computador/instrumentação , Software , Água
6.
Med Phys ; 33(10): 3921-30, 2006 Oct.
Artigo em Inglês | MEDLINE | ID: mdl-17089854

RESUMO

The aim of this study was to demonstrate how dosimetry with an amorphous silicon electronic portal imaging device (a-Si EPID) replaced film and ionization chamber measurements for routine pre-treatment dosimetry in our clinic. Furthermore, we described how EPID dosimetry was used to solve a clinical problem. IMRT prostate plans were delivered to a homogeneous slab phantom. EPID transit images were acquired for each segment. A previously developed in-house back-projection algorithm was used to reconstruct the dose distribution in the phantom mid-plane (intersecting the isocenter). Segment dose images were summed to obtain an EPID mid-plane dose image for each field. Fields were compared using profiles and in two dimensions with the y evaluation (criteria: 3%/3 mm). To quantify results, the average gamma (gamma avg), maximum gamma (gamma max), and the percentage of points with gamma < 1(P gamma < 1) were calculated within the 20% isodose line of each field. For 10 patient plans, all fields were measured with EPID and film at gantry set to 0 degrees. The film was located in the phantom coronal mid-plane (10 cm depth), and compared with the back-projected EPID mid-plane absolute dose. EPID and film measurements agreed well for all 50 fields, with (gamma avg) =0.16, (gamma max)=1.00, and (P gamma < 1)= 100%. Based on these results, film measurements were discontinued for verification of prostate IMRT plans. For 20 patient plans, the dose distribution was re-calculated with the phantom CT scan and delivered to the phantom with the original gantry angles. The planned isocenter dose (plan(iso)) was verified with the EPID (EPID(iso)) and an ionization chamber (IC(iso)). The average ratio, (EPID(iso)/IC(iso)), was 1.00 (0.01 SD). Both measurements were systematically lower than planned, with (EPID(iso)/plan(iso)) and (IC(iso)/plan(iso))=0.99 (0.01 SD). EPID mid-plane dose images for each field were also compared with the corresponding plane derived from the three dimensional (3D) dose grid calculated with the phantom CT scan. Comparisons of 100 fields yielded (gamma avg)=0.39, gamma max=2.52, and (P gamma < 1)=98.7%. Seven plans revealed under-dosage in individual fields ranging from 5% to 16%, occurring at small regions of overlapping segments or along the junction of abutting segments (tongue-and-groove side). Test fields were designed to simulate errors and gave similar results. The agreement was improved after adjusting an incorrectly set tongue-and-groove width parameter in the treatment planning system (TPS), reducing (gamma max) from 2.19 to 0.80 for the test field. Mid-plane dose distributions determined with the EPID were consistent with film measurements in a slab phantom for all IMRT fields. Isocenter doses of the total plan measured with an EPID and an ionization chamber also agreed. The EPID can therefore replace these dosimetry devices for field-by-field and isocenter IMRT pre-treatment verification. Systematic errors were detected using EPID dosimetry, resulting in the adjustment of a TPS parameter and alteration of two clinical patient plans. One set of EPID measurements (i.e., one open and transit image acquired for each segment of the plan) is sufficient to check each IMRT plan field-by-field and at the isocenter, making it a useful, efficient, and accurate dosimetric tool.


Assuntos
Neoplasias da Próstata/radioterapia , Radiometria/métodos , Planejamento da Radioterapia Assistida por Computador/métodos , Radioterapia de Intensidade Modulada/métodos , Algoritmos , Calibragem , Humanos , Íons , Masculino , Aceleradores de Partículas , Imagens de Fantasmas , Neoplasias da Próstata/patologia , Doses de Radiação , Dosagem Radioterapêutica , Espalhamento de Radiação
7.
Med Phys ; 33(7): 2448-51, 2006 Jul.
Artigo em Inglês | MEDLINE | ID: mdl-16898447

RESUMO

Many studies have reported dosimetric characteristics of amorphous silicon electronic portal imaging devices (EPIDs). Some studies ascribed a non-linear signal to gain ghosting and image lag. Other reports, however, state the effect is negligible. This study compares the signal-to-monitor unit (MU) ratio for three different brands of EPID systems. The signal was measured for a wide range of monitor units (5-1000), dose-rates, and beam energies. All EPIDs exhibited a relative under-response for beams of few MUs; giving 4 to 10% lower signal-to-MU ratios relative to that of 1000 MUs. This under-response is consistent with ghosting effects due to charge trapping.


Assuntos
Processamento de Imagem Assistida por Computador/métodos , Radiometria/métodos , Silício/química , Calibragem , Desenho de Equipamento , Humanos , Imagens de Fantasmas , Interpretação de Imagem Radiográfica Assistida por Computador , Dosagem Radioterapêutica , Reprodutibilidade dos Testes , Software , Fatores de Tempo
8.
Med Phys ; 31(11): 2989-95, 2004 Nov.
Artigo em Inglês | MEDLINE | ID: mdl-15587651

RESUMO

This study was carried out to determine the stability of the response of amorphous silicon (a-Si)-flat panel imagers for dosimetry applications. Measurements of the imager's response under reference conditions were performed on a regular basis for four detectors of the same manufacturer. We found that the ambient temperature influenced the dark-field, while the gain of the imager signal was unaffected. Therefore, temperature fluctuations were corrected for by applying a "dynamic" darkfield correction. This correction method also removed the influence of a small, irreversible increase of the dark-field current, which was equal to 0.5% of the dynamic range of the imager per year and was probably caused by mild radiation damage to the a-Si array. By applying a dynamic dark-field correction, excellent stability of the response over the entire panel of all imagers of 0.5% (1 SD) was obtained over an observation period up to 23 months. However, two imagers had to be replaced after several months. For one imager, an image segment stopped functioning, while the image quality of the other imager degraded significantly. We conclude that the tested a-Si EPIDs have a very stable response and are therefore well suited for dosimetry. We recommend, however, applying quality assurance tests dedicated to both imaging and dosimetry.


Assuntos
Radiometria/instrumentação , Planejamento da Radioterapia Assistida por Computador/instrumentação , Silício/efeitos da radiação , Calibragem , Relação Dose-Resposta à Radiação , Desenho de Equipamento , Análise de Falha de Equipamento , Garantia da Qualidade dos Cuidados de Saúde/métodos , Garantia da Qualidade dos Cuidados de Saúde/normas , Radiometria/normas , Dosagem Radioterapêutica , Planejamento da Radioterapia Assistida por Computador/métodos , Planejamento da Radioterapia Assistida por Computador/normas , Reprodutibilidade dos Testes , Sensibilidade e Especificidade , Temperatura
9.
Med Phys ; 31(2): 285-95, 2004 Feb.
Artigo em Inglês | MEDLINE | ID: mdl-15000614

RESUMO

The purpose of this study was to investigate the dose-response characteristics, including ghosting effects, of an amorphous silicon-based electronic portal imaging device (a-Si EPID) under clinical conditions. EPID measurements were performed using one prototype and two commercial a-Si detectors on two linear accelerators: one with 4 and 6 MV and the other with 8 and 18 MV x-ray beams. First, the EPID signal and ionization chamber measurements in a mini-phantom were compared to determine the amount of buildup required for EPID dosimetry. Subsequently, EPID signal characteristics were studied as a function of dose per pulse, pulse repetition frequency (PRF) and total dose, as well as the effects of ghosting. There was an over-response of the EPID signal compared to the ionization chamber of up to 18%, with no additional buildup layer over an air gap range of 10 to 60 cm. The addition of a 2.5 mm thick copper plate sufficiently reduced this over-response to within 1% at clinically relevant patient-detector air gaps (> 40 cm). The response of the EPIDs varied by up to 8% over a large range of dose per pulse values, PRF values and number of monitor units. The EPID response showed an under-response at shorter beam times due to ghosting effects, which depended on the number of exposure frames for a fixed frame acquisition rate. With an appropriate build-up layer and corrections for dose per pulse, PRF and ghosting, the variation in the a-Si EPID response can be reduced to well within +/- 1%.


Assuntos
Processamento de Imagem Assistida por Computador/métodos , Silício/química , Ar , Cobre , Relação Dose-Resposta à Radiação , Íons , Aceleradores de Partículas , Fótons , Poliestirenos , Radiometria , Sensibilidade e Especificidade , Fatores de Tempo