1.
Membranes (Basel) ; 8(3)2018 Sep 03.
Artigo em Inglês | MEDLINE | ID: mdl-30177638

RESUMO

This paper presents a novel method for transient and steady state mixed gas permeation measurements, using a quadrupole residual gas analyser for the on-line determination of the permeate composition. The on-line analysis provides sufficiently quick response times to follow even fast transient phenomena, enabling the unique determination of the diffusion coefficient of the individual gases in a gas mixture. Following earlier work, the method is further optimised for higher gas pressures, using a thin film composite and a thick dense styrene-butadiene-styrene (SBS) block copolymer membrane. Finally, the method is used to calculate the CO2/CH4 mixed gas diffusion coefficients of the spirobisfluorene-based polymer of intrinsic microporosity, PIM-SBF-1. It is shown that the modest pressure dependence of the PIM-SBF-1 permeability can be ascribed to a much stronger pressure dependence of the diffusion coefficient, which partially compensates the decreasing solubility of CO2 with increasing pressure, typical for the strong sorption behaviour in PIMs. The characteristics of the instrument are discussed and suggestions are given for even more versatile measurements under stepwise increasing pressure conditions. This is the first report on mixed gas diffusion coefficients at different pressures in a polymer of intrinsic microporosity.