Rev Sci Instrum ; 88(10): 105103, 2017 Oct.
Artigo em Inglês | MEDLINE | ID: mdl-29092508

RESUMO

A combined optical coherence tomography (OCT)-Raman probe was designed and built into a spectral domain OCT head, and its performance was evaluated and compared to the most common Raman probe setups, based on a fiber bundle and confocal free space optics. Due to the use of the full field of view of an OCT scanning lens, the combined probe has a superior performance within maximum permissible exposure limits, compared to the other two probes. Skin Raman spectra, recorded in vivo, further prove the feasibility of the OCT-Raman probe for the future in vivo clinical applications in skin cancer screening.


Assuntos
Melanoma/diagnóstico por imagem , Neoplasias Cutâneas/diagnóstico por imagem , Tomografia de Coerência Óptica , Detecção Precoce de Câncer , Humanos , Melanoma Maligno Cutâneo