1.
Appl Opt ; 41(7): 1315-22, 2002 Mar 01.
Artigo em Inglês | MEDLINE | ID: mdl-11900009

RESUMO

We have developed a low-coherence interferometer system used for the simultaneous measurement of refractive index n and thickness t of transparent plates. Both the phase index n(p) and group index n(g) can be determined automatically in a wide thickness range of from 10 microm to a few millimeters. Two unique techniques are presented to measure n(p), n(g), and t simultaneously. One allows us to determine n(p), n(g), and t accurately by using a special sample holder, in which the measurement accuracy is 0.3% for the thickness t above 0.1 mm. In the other technique the chromatic dispersion delta n of index is approximately expressed as a function of (n(p) - 1) on the basis of measured values of n(p) and n(g) for a variety of materials, and then the simultaneous measurement is performed with a normal sample holder. In addition, a measurement accuracy of less than 1% is achieved even when the sample is as thin as 20 microm. The measurement time is also 3 min or more.


Assuntos
Interferometria/instrumentação , Refratometria , Calibragem , Modelos Teóricos
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...