Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros










Base de dados
Intervalo de ano de publicação
1.
Indian J Med Microbiol ; 50: 100663, 2024 Jul 05.
Artigo em Inglês | MEDLINE | ID: mdl-38964501

RESUMO

PURPOSE: Candida albicans is the second most common cause of candidemia in Malaysia. The Clinical and Laboratory Standards Institute (CLSI) broth microdilution method is the gold standard for determining its minimum inhibitory concentration (MIC); however, it is laborious and time-consuming. This study was conducted to evaluate the usefulness of alternative methods, namely Sensititre YeastOne (SYO), VITEK 2 system, and E-test for determining the MIC of clinical C. albicans isolates. MATERIALS AND METHODS: The susceptibilities of 95 C. albicans isolates were compared between SYO, VITEK 2 system, and E-test with CLSI broth microdilution method. The categorical agreement (CA), essential agreement (EA), very major errors (VME), major errors (ME) and minor errors (MiE) were calculated. RESULTS: Our finding showed the CA varied for SYO from 96.8% to 100%, while the EA ranged from 91.6% to 100%. The SYO method showed 1.1% of VME and ME, and up to 3.2% of MiE. Next, the CA and EA ranges for the VITEK 2 system were 97.8%-100% and 23.2%-100%, respectively. In the VITEK 2 technique, 1.1% of VME were found. For the E-test, the CA varied from 83.2% to 100% while the EA ranged from 64.2% to 98.9%. The E-test method showed 1.1% of VME and up to 16.8% of MiE. CONCLUSIONS: In conclusion, SYO and VITEK 2 (except flucytosine) could be potential alternatives to the CLSI broth microdilution method in determining the MIC of C. albicans.

SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...