1.
Anesth Analg ; 124(1): 95-103, 2017 01.
Artigo em Inglês | MEDLINE | ID: mdl-27992386

RESUMO

BACKGROUND: Within the context of automating neonatal oxygen therapy, this article describes the transformation of an idea verified by a computer model into a device actuated by a computer model. Computer modeling of an entire neonatal oxygen therapy system can facilitate the development of closed-loop control algorithms by providing a verification platform and speeding up algorithm development. METHODS: In this article, we present a method of mathematically modeling the system's components: the oxygen transport within the patient, the oxygen blender, the controller, and the pulse oximeter. Furthermore, within the constraints of engineering a product, an idealized model of the neonatal oxygen transport component may be integrated effectively into the control algorithm of a device, referred to as the adaptive model. Manual and closed-loop oxygen therapy performance were defined in this article by 3 criteria in the following order of importance: percent duration of SpO2 spent in normoxemia (target SpO2 ± 2.5%), hypoxemia (less than normoxemia), and hyperoxemia (more than normoxemia); number of 60-second periods <85% SpO2 and >95% SpO2; and number of manual adjustments. RESULTS: Results from a clinical evaluation that compared the performance of 3 closed-loop control algorithms (state machine, proportional-integral-differential, and adaptive model) with manual oxygen therapy on 7 low-birth-weight ventilated preterm babies, are presented. Compared with manual therapy, all closed-loop control algorithms significantly increased the patients' duration in normoxemia and reduced hyperoxemia (P < 0.05). The number of manual adjustments was also significantly reduced by all of the closed-loop control algorithms (P < 0.05). CONCLUSIONS: Although the performance of the 3 control algorithms was equivalent, it is suggested that the adaptive model, with its ease of use, may have the best utility.


Assuntos
Algoritmos , Simulação por Computador , Hipóxia/terapia , Modelos Biológicos , Oxigenoterapia/métodos , Respiração Artificial/métodos , Terapia Assistida por Computador/métodos , Biomarcadores/sangue , Feminino , Humanos , Hiperóxia/sangue , Hiperóxia/diagnóstico , Hiperóxia/etiologia , Hipóxia/sangue , Hipóxia/diagnóstico , Hipóxia/etiologia , Recém-Nascido de Baixo Peso , Recém-Nascido , Recém-Nascido Prematuro , Masculino , Oximetria , Oxigênio/sangue , Oxigenoterapia/efeitos adversos , Respiração Artificial/efeitos adversos , Fatores de Tempo
2.
Artigo em Inglês | MEDLINE | ID: mdl-19963561

RESUMO

Neonates with under developed lungs often require oxygen therapy. During the course of oxygen therapy, elevated levels of blood oxygenation, hyperoxemia, must be avoided or the risk of chronic lung disease or retinal damage is increased. Low levels of blood oxygen, hypoxemia, may lead to permanent brain tissue damage and, in some cases, mortality. A closed loop controller that automatically administers oxygen therapy using 3 algorithms - state machine, adaptive model, and proportional integral derivative (PID) - is applied to 7 ventilated low birth weight neonates and compared to manual oxygen therapy. All 3 automatic control algorithms demonstrated their ability to improve manual oxygen therapy by increasing periods of normoxemia and reducing the need for manual FiO(2) adjustments. Of the three control algorithms, the adaptive model showed the best performance with 0.25 manual adjustments per hour and 73% time spent within target +/- 3% SpO(2).


Assuntos
Hipóxia/terapia , Oxigênio/sangue , Oxigênio/uso terapêutico , Respiração Artificial/métodos , Algoritmos , Simulação por Computador , Desenho de Equipamento , Feminino , Humanos , Recém-Nascido de Baixo Peso , Recém-Nascido , Masculino , Oximetria , Oxigenoterapia , Ventilação Pulmonar , Testes de Função Respiratória