Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros










Base de dados
Intervalo de ano de publicação
1.
Artigo em Japonês | MEDLINE | ID: mdl-32074527

RESUMO

PURPOSE: In this study, we evaluated the stability and reliability of absorbed dose-to-water for an HDR 192Ir sandwich setup phantom method by comparing measurements with absorbed dose-to-water determination based on the AAPM TG-43 protocol. METHODS: The sandwich setup phantom was designed with a dedicated device for two ion chamber measurements of absorbed dose-to-water for a mHDR-v2r 192Ir brachytherapy source is presented. To test the reliability of sandwich setup phantom of measurements with absorbed dose-to-water, we were compared with values based on AAPM TG-43 protocol and evaluated temporal variations of the measurement, intra-rater reliability. RESULTS: The measured doses at sandwich setup phantom agreed within 1.0% with AAPM TG-43 protocol. In all measurement fractions, the temporal variations of measurement value were less than 1.0%, and the intra-rater reliability were 0.94% or more. CONCLUSIONS: The measurement value obtained by the absorbed dose-towater had good reliability, and sandwich setup phantom is potentially useful and convenient for daily dose management of 192Ir sources in clinics.


Assuntos
Braquiterapia , Imagens de Fantasmas , Água , Radioisótopos de Irídio , Método de Monte Carlo , Radiometria , Dosagem Radioterapêutica , Reprodutibilidade dos Testes
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...