Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 3 de 3
Filtrar
Mais filtros











Base de dados
Intervalo de ano de publicação
1.
Rev Mal Respir ; 31(4): 312-22, 2014 Apr.
Artigo em Francês | MEDLINE | ID: mdl-24750951

RESUMO

Non-invasive ventilation (NIV) is recognised as an effective treatment for chronic hypercapnic respiratory failure. Monitoring NIV during sleep may be preferable to daytime assessment. This paper reports the findings of an international consensus group, which systematically analysed nocturnal polygraphic or polysomnographic tracings recorded with either volume-cycled or pressure-cycled ventilators. A systematic description of nocturnal respiratory events, which occur during NIV, is proposed: leaks, obstruction at different levels of the upper airway (glottis and/or pharynx), with or without decrease of respiratory drive and asynchrony.

2.
Thorax ; 67(6): 546-52, 2012 Jun.
Artigo em Inglês | MEDLINE | ID: mdl-20971982

RESUMO

Non-invasive ventilation (NIV) is recognised as an effective treatment for chronic hypercapnic respiratory failure. Monitoring NIV during sleep may be preferable to daytime assessment. This paper reports the findings of an international consensus group which systematically analysed nocturnal polygraphic or polysomnographic tracings recorded with either volume-cycled or pressure-cycled ventilators. A systematic description of nocturnal respiratory events which occur during NIV is proposed: leaks, obstruction at different levels of the upper airway (glottis and/or pharynx), with or without decrease of respiratory drive and asynchrony.


Assuntos
Polissonografia , Respiração com Pressão Positiva , Síndromes da Apneia do Sono/fisiopatologia , Síndromes da Apneia do Sono/terapia , Conferências de Consenso como Assunto , Humanos , Monitorização Fisiológica/métodos , Guias de Prática Clínica como Assunto , Testes de Função Respiratória , Insuficiência Respiratória/terapia , Mecânica Respiratória , Sono , Síndromes da Apneia do Sono/diagnóstico , Resultado do Tratamento
3.
Eur Respir J ; 19(1): 108-12, 2002 Jan.
Artigo em Inglês | MEDLINE | ID: mdl-11843308

RESUMO

A bench study using an artificial lung model was performed to evaluate the snoring detection sensitivity of six (commercially available) auto-nasal continuous positive airway pressure (NCPAP) devices. Snoring was simulated by a loudspeaker connected to the lung model and abruptly activated during 1 s of each inspiratory period to induce pressure oscillation. The oscillation frequencies chosen were 30, 60, 90, and 120 Hz. For each frequency, the amplitude of the pressure oscillation produced by the loudspeaker was adjusted to find the threshold at which the auto-nCPAP devices detected snoring. Differences in pressure-amplitude thresholds of up to three-fold were found across auto-nCPAP devices. A randomized clinical study to compare the effects of the least sensitive (Virtuoso LX; Respironics, Nantes, France) and one of the most sensitive, (Goodknight 418A; Malinckrodt, Nancy, France) devices, in two groups of six patients with obstructive sleep apnoea syndrome was then conducted. Goodknight 418A was more sensitive than Virtuoso LX for detecting snoring (mean +/- SD 92 +/- 11% versus 50 +/- 39% respectively, p = 0.03). To conclude, striking differences exist between auto-nasal continuous positive airway pressure devices in sensitivity for detecting snoring.


Assuntos
Nariz/fisiologia , Respiração com Pressão Positiva , Ronco/diagnóstico , Humanos , Pessoa de Meia-Idade , Modelos Teóricos , Apneia Obstrutiva do Sono/diagnóstico
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA