1.
Artigo em Inglês | MEDLINE | ID: mdl-26186723

RESUMO

The analysis of human plasma free amino acids is important for diagnosing the health of individuals, because their concentrations are known to vary with various diseases. The development of valid, reliable, and high-throughput analytical methods for amino acids analysis is an essential requirement in clinical applications. In the present study, we have developed an automated precolumn derivatization amino acid analytical method based on high-performance liquid chromatography/electrospray ionization mass spectrometry (so-called UF-Amino Station). This method enabled the separation of at least 38 types of physiological amino acids within 8min, and the interval time between injections was 12min. We also validated this method for 21 major types of free amino acids in human plasma samples. The results of the specificity, linearity, accuracy, repeatability, intermediate precision, reproducibility, limits of detections, lower limits of quantification, carry over, and sample solution stability were sufficient to allow for the measurement of amino acids in human plasma samples. Our developed method should be suitable for use in clinical fields.


Assuntos
Aminoácidos/sangue , Cromatografia Líquida de Alta Pressão/métodos , Espectrometria de Massas por Ionização por Electrospray/métodos , Humanos , Espectrometria de Massas por Ionização por Electrospray/instrumentação