Artigo em Inglês | MEDLINE | ID: mdl-38083409

RESUMO

We obtain and compare the non-pulsating part of reflective Photoplethysmogram (PPG) measurements in a porcine skin phantom and a wearable device prototype with Monte Carlo simulations and analyse the received signal. In particular, we investigate typical PPG wavelengths at 520, 637 and 940 nm and source-detector distances between 1.5 and 8.0 mm. We detail the phantom's optical parameters, the wearable device design, and the simulation setup. Monte Carlo simulations were using layer-based and voxel-based structures. Pattern of the detected photon weights showed comparable trends. PPG signal, differential pathlength factor (DPF), mean maximum penetration depth, and signal level showed dependencies on the source-detector distance d for all wavelengths. We demonstrate the signal dependence of emitter and detection angles, which is of interest for the development of wearables.


Assuntos
Fótons , Método de Monte Carlo , Simulação por Computador
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...