Patient dose reduction in computed tomography (CT) always results in a trade off between radiation exposure and image quality. There are few reports that estimate the relationship between image quality and X-ray exposure in CT examinations as one optimal index. The purpose of this study was to determine the optimal parameter settings enabling a low radiation exposure without compromising image quality using a particular 4-row multislice CT (MSCT) scanner (Aquilion VZ 4-slice CT scanner, Toshiba Medical Systems Corporation, Otawara, Tochigi, Japan). Normalized dose divided by image noise for helical pitches (nDNR: normalized dose to noise ratio) were calculated in consideration of beam collimation and tube current-time product. Optimal tube current-time product was calculated using the nDNR for the helical pitches based on user-defined standards of quality of the CT image. As a result, the nDNR proved to be well-supported to decrease the patient exposure in various exposure conditions of MSCT scans; however, the dose and image noise did not show a linear relation to the helical pitch. In conclusion, nDNR can be applied to patient dose reduction while keeping an acceptable image quality using a particular 4-row MSCT scanner.