1.
Mater Sci Eng C Mater Biol Appl ; 33(2): 870-4, 2013 Mar 01.
Artigo em Inglês | MEDLINE | ID: mdl-25427500

RESUMO

A new ion-selective electrode for Sm(3+) ion is described based on the incorporation of N,N',N"-tris(4-pyridyl)trimesic amide (TPTA) in a poly(vinylchloride) (PVC) matrix. The membrane sensor comprises nitrobenzene (NB) as a plasticizer, and oleic acid (OA) as an anionic additive. The sensor with the optimized composition shows a Nernstian potential response of 19.8 ± 0.5 mV decade(-1) over a wide concentration range of 1.0 × 10(-2) and 1 × 10(-6)mol L(-1), with a lower detection limit of 4.7 × 10(-7)mol L(-1) and satisfactor applicable pH range of 3.6-9.2. Having a short response time of less than 10s and a very good selectivity towards the Sm(3+) over a wide variety of interfering cations (e.g. alkali, alkaline earth, transition and heavy metal ions) the sensor seemed to be a promising analytical tool for determination of the Sm(3+). Hence, it was used as an indicator electrode in the potentiometric titration of samarium ion with EDTA. It was also applied to the direct samarium recovery in binary mixtures.


Assuntos
Eletrodos Seletivos de Íons , Membranas Artificiais , Cloreto de Polivinila/química , Potenciometria/instrumentação , Samário/análise , Calibragem , Concentração de Íons de Hidrogênio , Limite de Detecção , Samário/química
SELEÇÃO DE REFERÊNCIAS