1.
Med Eng Phys ; 31(10): 1337-42, 2009 Dec.
Artigo em Inglês | MEDLINE | ID: mdl-19819750

RESUMO

To determine the reliability, stability, validity and precision of a stereophotogrammetry (SP) system for use in quantifying the complex three-dimensional structure of the human torso, we performed assessments of the system using images of geometric solids and a human-form mannequin. Analysis of geometric solids revealed excellent intra- and interrater reliability of the system for linear, surface area and volume measurements (r>0.99, P<0.001). Overall, no significant difference was found between SP and manual measurements (F=4.23, P>0.06). The system exhibited excellent stability in images of the mannequin over time (r>0.99). The limit of precision (error>5%) of the system to detect objects on the surface of the mannequin was estimated at an object size of 23.5cm(2) for surface area and 32mL for volume. These results demonstrate the capability of SP of the torso to be used as a reliable, stable and valid measure of torso morphology to be applied as a clinical outcome tool in studies of bony and soft tissue pathologies such as scoliosis, rib deformities, obesity or edema.


Assuntos
Fotogrametria/métodos , Escoliose/patologia , Antropometria , Desenho de Equipamento , Feminino , Humanos , Processamento de Imagem Assistida por Computador , Imageamento Tridimensional/métodos , Modelos Anatômicos , Análise de Regressão , Reprodutibilidade dos Testes , Software , Propriedades de Superfície
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...