Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros










Base de dados
Intervalo de ano de publicação
1.
Talanta ; 218: 121118, 2020 Oct 01.
Artigo em Inglês | MEDLINE | ID: mdl-32797876

RESUMO

A simple cost-effective moving drop device, with solenoid valves control for programmable liquid handling, was developed for electrokinetic injection in capillary zone electrophoresis (CZE). With a CZE of 50-µm bore fused silica capillary tube and contactless conductivity detector (C4D), mixed anions (Cl-, NO3-, and SO42- as the model) solution was injected at the ground side. Simultaneous quantitative chemical analysis can be achieved. A linear relationship of concentration and the peak height was achieved in the range of 0.5-10 mg L-1 for each anion, with LOD and LOQ being 0.02 mg L-1 and 0.5 mg L-1, respectively. The MVD-CZE system allows continuous operation with a sample throughput of 40 samples/hour. A real sample application was demonstrated for air samples, with one drop (25 µL) collected from a midget bubbler based air sampling unit. The recoveries were found to be 74.4-115.0 %. The developed device was also preliminarily applied for the injection of mixed cations (NH4+, Na+, K+, Ca2+, and Mg2+) at the high voltage side end with optimistic results.