RESUMO
BACKGROUND: The etiology of transcatheter aortic valve (TAV) degeneration is poorly understood, particularly noncalcific mechanisms. OBJECTIVES: The authors sought to investigate noncalcific and calcific mechanisms of TAV degeneration and evaluate their impact on leaflet function by bench testing, imaging, and histology. METHODS: TAV explants were obtained from the EXPLANT THV registry and clinical institutions. Hydrodynamic assessment was performed using a heart valve pulse duplicator system under physiological conditions. Micro-computed tomography, high-resolution photography, high speed video, and hematoxylin and eosin staining were used to evaluate the morphological appearance, leaflet kinematics, and calcium burden of TAVs. RESULTS: A total of 14 explants were evaluated: 10 self-expanding CoreValve/Evolut TAVs (Medtronic), 3 balloon-expandable SAPIEN 3 TAVs (Edwards Lifesciences), and 1 mechanically expandable Lotus TAV (Boston Scientific). The median patient age at explantation was 73.0 years (Q1-Q3: 64.5-80.0 years), with a time to explantation of 4 years 1 month (1 year 5 months to 4 years 11 months). Six TAV explants were found to have leaflet calcification (162.4 mm3; 58.8-603.0 mm3), and 8 had no calcification detectable by micro-computed tomography and histology. All samples had impaired leaflet kinematics. There was no significant difference in the hydrodynamic mean gradient between calcified (47.2 mm Hg; 26.6-74.1 mm Hg) and noncalcified (27.6 mm Hg; 15.2-36.7 mm Hg; P = 0.28) TAVs. Leaflet calcification had a weak but nonsignificant association with the hydrodynamic mean gradient (r = 0.42; P = 0.14). CONCLUSIONS: TAV function can be severely impacted by noncalcific and calcific mechanisms of tissue degeneration. Importantly, functional stenosis can occur in TAVs in the absence of obvious and significant calcification.