Artif Organs ; 40(12): 1113-1120, 2016 Dec.
Artigo em Inglês | MEDLINE | ID: mdl-27230977

RESUMO

New left ventricular assist devices (LVADs) offer both important advantages and potential hazards. VAD development requires better and expeditious ways to identify these advantages and hazards. We validated in an isolated working heart the hemodynamic performance of an intraventricular LVAD and investigated how its outflow cannula interacted with the aortic valve. Hearts from six pigs were explanted and connected to an isolated working heart setup. A miniaturized LVAD was implanted within the left ventricle (tMVAD, HeartWare Inc., Miami Lakes, FL, USA). In four experiments blood was used to investigate hemodynamics under various loading conditions. In two experiments crystalloid perfusate was used, allowing visualization of the outflow cannula within the aortic valve. In all hearts the transapical miniaturized ventricular assist device (tMVAD) implantation was successful. In the blood experiments hemodynamics similar to those observed clinically were achieved. Pump speeds ranged from 9 to 22 krpm with a maximum of 7.6 L/min against a pressure difference between ventricle and aorta of ∼50 mm Hg. With crystalloid perfusate, central positioning of the outflow cannula in the aortic root was observed during full and partial support. With decreasing aortic pressures the cannula tended to drift toward the aortic root wall. The tMVAD could unload the ventricle similarly to LVADs under conventional cannulation. Aortic pressure influenced central positioning of the outflow cannula in the aortic root. The isolated heart is a simple, accessible evaluation platform unaffected by complex reactions within a whole, living animal. This platform allowed detection and visualization of potential hazards.


Assuntos
Valva Aórtica/fisiopatologia , Ventrículos do Coração/fisiopatologia , Ventrículos do Coração/cirurgia , Coração Auxiliar , Animais , Hemodinâmica , Miniaturização , Desenho de Prótese , Implantação de Prótese , Suínos , Função Ventricular Esquerda
2.
ASAIO J ; 62(1): 15-9, 2016.
Artigo em Inglês | MEDLINE | ID: mdl-26479467

RESUMO

Controller algorithms are an important feature for assessment of ventricular assist device performance. Flow estimation is one algorithm implemented in the HeartWare continuous-flow ventricular assist device pump system. This parameter estimates flow passing through the pump and is calculated using speed, current, and hematocrit. In vitro and in vivo studies were conducted to assess the algorithm accuracy. During in vitro testing, three pumps were tested in four water-glycerol solutions at 37°C with viscosities equivalent to hematocrits of 20, 30, 40, and 50%. By using a linear regression model, a correlation coefficient of >0.94 was observed between measured and estimated flow for all conditions. In vivo studies (n = 9) were conducted in an ovine model where a reference flow probe was placed on the outflow graft and speed was adjusted from 1,800 to 4,000 revolutions per minute. During in vivo experiments, estimated pump flow (mean, minimum, and maximum) was compared with measured pump flow. The best-fit linear regression equation for the data is y = 0.96x + 0.54, r = 0.92. In addition, waveform fidelity was high (r > 0.96) in normal (i.e., nonsuction) cases where flow pulsatility was >2 L/min. The flow estimation algorithm demonstrated strong agreement with measured flow, both when analyzing average waveform magnitude and fidelity.


Assuntos
Coração Auxiliar/normas , Fluxo Pulsátil , Algoritmos , Animais , Velocidade do Fluxo Sanguíneo , Desenho de Equipamento , Modelos Cardiovasculares , Ovinos