1.
Anesteziol Reanimatol ; (6): 6-8, 1990.
Artigo em Russo | MEDLINE | ID: mdl-2075937

RESUMO

The analysis of 65 cases of cardiopulmonary bypass surgery using systems distinct in the type of the oxygenator used (contact, membrane) and the degree of thromboresistance of blood-contacting surfaces has demonstrated that the use of heparin-covered membrane blood oxygenators and systems allows of a considerable reduction in blood trauma during perfusion and a 5-6-fold decrease of heparin doses employed for extracorporeal membrane blood oxygenation.


Assuntos
Ponte Cardiopulmonar , Heparina , Oxigenadores de Membrana , Adulto , Feminino , Humanos , Masculino , Pessoa de Meia-Idade