J Arthroplasty ; 24(1): 65-70, 2009 Jan.
Artigo em Inglês | MEDLINE | ID: mdl-18617364

RESUMO

We investigated 55 hips of 53 patients with dysplastic hips. Individual computed tomography data were used in the manufacturing of cementless custom-made stems made of Ti-6Al-4V. The proximal one third was coated using porous coating covered with hydroxyapatite coating. The average age at surgery was 60 years and the average follow-up was 7 years. Five patients (9%) complained of postoperative thigh pain. According to Engh's radiologic classification system, there was bone-ingrown fixation in all hips. The cortical index and canal flare index were lower, and stem diameter was greater in the 11 hips (20%) with severe stress shielding than in the remaining 44 hips. Although the results obtained with this custom-made stem system for dysplastic hips were excellent, stress shielding is still an issue.


Assuntos
Artroplastia de Quadril/instrumentação , Doenças do Desenvolvimento Ósseo/cirurgia , Desenho Assistido por Computador , Prótese de Quadril , Desenho de Prótese , Adulto , Idoso , Fenômenos Biomecânicos/fisiologia , Doenças do Desenvolvimento Ósseo/etnologia , Durapatita , Feminino , Seguimentos , Articulação do Quadril/diagnóstico por imagem , Articulação do Quadril/fisiologia , Articulação do Quadril/cirurgia , Humanos , Japão , Estudos Longitudinais , Masculino , Pessoa de Meia-Idade , Radiografia , Estresse Mecânico , Resultado do Tratamento