Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros










Base de dados
Intervalo de ano de publicação
1.
Healthc Technol Lett ; 11(2-3): 59-66, 2024.
Artigo em Inglês | MEDLINE | ID: mdl-38638487

RESUMO

This work presents a proof-of-concept of a robotic-driven intra-operative scanner designed for knee cartilage lesion repair, part of a system for direct in vivo bioprinting. The proposed system is based on a photogrammetric pipeline, which reconstructs the cartilage and lesion surfaces from sets of photographs acquired by a robotic-handled endoscope, and produces 3D grafts for further printing path planning. A validation on a synthetic phantom is presented, showing that, despite the cartilage smooth and featureless surface, the current prototype can accurately reconstruct osteochondral lesions and their surroundings with mean error values of 0.199 ± 0.096 mm but with noticeable concentration on areas with poor lighting or low photographic coverage. The system can also accurately generate grafts for bioprinting, although with a slight tendency to underestimate the actual lesion sizes, producing grafts with coverage errors of -12.2 ± 3.7, -7.9 ± 4.9, and -15.2 ± 3.4% for the medio-lateral, antero-posterior, and craneo-caudal directions, respectively. Improvements in lighting and acquisition for enhancing reconstruction accuracy are planned as future work, as well as integration into a complete bioprinting pipeline and validation with ex vivo phantoms.

SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...