1.
Am J Nephrol ; 20(5): 364-8, 2000.
Artigo em Inglês | MEDLINE | ID: mdl-11092992

RESUMO

We performed a crossover study to compare the effects of different dialysis membranes on 20 patients with frequent dialyser clotting and requiring > or = 5,000 units of heparin per dialysis session. Low-flux dialysers are C15NL (cellulose - Terumo) and E15NL (vitamin-E-coated - Terumo) while high-flux dialysers were F60 (polysulphone) and EE15NL (vitamin-E-coated - Terumo). Ten patients underwent dialysis for 2 months with C15NL then switched to E15NL for 2 months. Similarly, the other 10 patients were started on the high-flux dialyser F60 and then switched over to EE15NL for 2 months. The following parameters were measured at the beginning of the study, 2 weeks, 1 month and then at 2 months: hemoglobin, prothrombin time (PT), partial thromboplastin time (PTT), fibrinogen, protein C, protein S, antithrombin III (ATIII) and factor 12 activity. Dialyser clotting, heparin and erythropoietin requirements were assessed during each dialysis session. There was a significant reduction in clotting with E15NL in comparison to C15NL (22.8 +/- 17 and 44.1 +/- 22.8 (p = 0.0233), respectively). Similarly, heparin requirements were less in the vitamin-E-coated (E15NL) dialysers, 4, 754 +/- 1,427 vs. 6,011 +/- 856 units (p = 0.0281) and erythropoietin usage was also significantly reduced, 4,630 +/- 2,620 vs. 7,850 +/- 4,069 units (p = 0.049). There was a significant increase in hemoglobin with E15NL compared to C15NL, 115 +/- 10.4 vs. 108 +/- 13.1 (p = 0.0343). When the high-flux dialysers were compared there was a tendency towards less dialyser clotting with the EE15NL compared to F60, though this did not achieve statistical significance (p = 0.0561). We could not demonstrate any significant changes between the different dialysers with regards to PT, PTT, fibrinogen factor 12 activity, protein C, protein S and ATIII. In conclusion, we have shown that the use of vitamin-E-modified dialysers is associated with less clotting in patients with persistent clotting problems. In addition, this was associated with less heparin and erythropoietin requirements.


Assuntos
Anticoagulantes/administração & dosagem , Transtornos da Coagulação Sanguínea/prevenção & controle , Materiais Revestidos Biocompatíveis , Eritropoetina/administração & dosagem , Heparina/administração & dosagem , Membranas Artificiais , Diálise Renal/efeitos adversos , Vitamina E/uso terapêutico , Adulto , Anticoagulantes/uso terapêutico , Estudos Cross-Over , Relação Dose-Resposta a Droga , Eritropoetina/uso terapêutico , Feminino , Heparina/uso terapêutico , Humanos , Masculino , Pessoa de Meia-Idade , Resultado do Tratamento
2.
Saudi J Kidney Dis Transpl ; 11(4): 543-7, 2000.
Artigo em Inglês | MEDLINE | ID: mdl-18209342

RESUMO

We performed a crossover study to compare the effect of vitamin E-modified dialyzers on acute intra-dialytic symptoms, with other membranes. Twenty patients on hemodialysis were studied. They were divided into two equal groups of low-flux (C15NL, E15NL) and high-flux (F60, EE15NL) membrane dialyzers. Within each group, a vitamin E-modified dialyzer was compared with another dialyzer in a crossover design over a two-month period. All study patients were seen during each dialysis session by a physician and the occurrence of intra-dialytic symptoms were recorded. There was a significant overall improvement in the incidence of acute intra-dialytic symptoms with the use of vitamin E-modified dialyzers as compared with the other membranes. This effect was more for cuprophane than polysulfone. The occurrence of hypotensive episodes did not differ. Our study indicates that we can achieve a reduction in the incidence of acute intra-dialytic symptoms with the use of vitamin E-modified membrane as compared to cuprophane and polysulfone.

SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...