Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros










Base de dados
Intervalo de ano de publicação
1.
Magn Reson Imaging ; 9(3): 395-408, 1991.
Artigo em Inglês | MEDLINE | ID: mdl-1881259

RESUMO

The radio-frequency 'birdcage resonator' has found wide use in MRI/MRS for its field homogeneity and signal-noise characteristics. This paper presents a general analysis, derived from lumped element transmission line theory, of the electrical behavior of unloaded, N-column birdcage resonators applicable to several versions of the basic design including low-pass and high-pass coils. Analytic expressions and computer results are presented for both types of coil describing resonant frequencies, input and characteristic impedances, dispersion relations, pass-bands, resonant peak bandwidth and Q. Theoretical expressions for normalized resonant frequency difference ratios independent of element values and resonator geometry have been developed for generic low- and high-pass coils. Experimental measurements of resonant frequencies were made for six coils, and the average agreement with theoretical predictions was approximately 4%.


Assuntos
Imageamento por Ressonância Magnética/instrumentação , Espectroscopia de Ressonância Magnética/instrumentação , Imageamento por Ressonância Magnética/métodos , Espectroscopia de Ressonância Magnética/métodos , Fenômenos Físicos , Física
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...