1.
Neuroimage ; 224: 117355, 2021 01 01.
Artigo em Inglês | MEDLINE | ID: mdl-32916290

RESUMO

PURPOSE: Multichannel Transcranial Magnetic Stimulation (mTMS) arrays enable multiple sites to be stimulated simultaneously or sequentially under electronic control without moving the system's stimulation coils. Here, we build and characterize the performance of a novel modular 3-axis TMS coil that can be utilized as a unit element in large-scale multichannel TMS arrays. METHODS: We determined the basic physical characteristics of the 3-axis TMS coil x-, y- and z-elements using a custom 2-channel programmable stimulator prototype. We mapped the temporal rate-of-change of the induced magnetic field (dB/dt) on a 2D plane parallel to the coil surface (including an extended line for full spatial coverage) and compared those values with predictions from magnetic field simulations. Temperature measurements were carried out to assess the incorporated air-cooling method. We measured the mutual and self-inductances of the x/y/z-elements to assess coupling between them. Additionally, we measured and calculated the coupling between z-elements in the array configuration. Finally, we performed electric field simulations to evaluate the stimulation intensity and focality of the coil and compared the results to conventional TMS coils as well as demonstrated suitability of the 3-axis coil for a multichannel array configuration. RESULTS: The experimentally obtained dB/dt values validated the computational model of the 3-axis coil and therefore confirmed that both the coil and stimulator system are operating as intended. The air-cooling system was effective for brief high-frequency pulse trains and extended single- and paired-pulse TMS protocols. The electromagnetic simulations suggested that an array of the 3-axis coils would have comparable stimulation intensity to conventional TMS coils, therefore enabling clearly suprathreshold stimulation of the human brain. The recorded coil coupling between the x/y/z-elements was < 1% and the maximal coupling between z-elements in the array configuration was 1.8% and 3.4% for the measured and calculated values, respectively. CONCLUSION: We presented a 3-axis coil intended for multichannel TMS arrays. The electromagnetic measurements and simulations verified that the coil fabrication met the desired specifications and that the inductive coupling between the elements was negligible. The air-cooled 3-axis TMS coil appears suitable to be used as an element in multichannel TMS arrays.


Assuntos
Encéfalo/fisiologia , Campos Eletromagnéticos , Estimulação Magnética Transcraniana , Simulação por Computador , Cabeça/fisiologia , Frequência Cardíaca/fisiologia , Humanos , Estimulação Magnética Transcraniana/métodos
2.
Rev Sci Instrum ; 88(5): 055110, 2017 May.
Artigo em Inglês | MEDLINE | ID: mdl-28571426

RESUMO

Versatile controllers for accurate, fast, and real-time synchronized acquisition of large-scale data are useful in many areas of science, engineering, and technology. Here, we describe the development of a controller software based on a technique called queued state machine for controlling the data acquisition (DAQ) hardware, continuously acquiring a large amount of data synchronized across a large number of channels (>400) at a fast rate (up to 20 kHz/channel) in real time, and interfacing with applications for real-time data analysis and display of electrophysiological data. This DAQ controller was developed specifically for a 384-channel pediatric whole-head magnetoencephalography (MEG) system, but its architecture is useful for wide applications. This controller running in a LabVIEW environment interfaces with microprocessors in the MEG sensor electronics to control their real-time operation. It also interfaces with a real-time MEG analysis software via transmission control protocol/internet protocol, to control the synchronous acquisition and transfer of the data in real time from >400 channels to acquisition and analysis workstations. The successful implementation of this controller for an MEG system with a large number of channels demonstrates the feasibility of employing the present architecture in several other applications.

3.
Rev Sci Instrum ; 87(9): 094301, 2016 Sep.
Artigo em Inglês | MEDLINE | ID: mdl-27782541

RESUMO

We developed a 375-channel, whole-head magnetoencephalography (MEG) system ("BabyMEG") for studying the electrophysiological development of human brain during the first years of life. The helmet accommodates heads up to 95% of 36-month old boys in the USA. The unique two-layer sensor array consists of: (1) 270 magnetometers (10 mm diameter, ∼15 mm coil-to-coil spacing) in the inner layer, (2) thirty-five three-axis magnetometers (20 mm × 20 mm) in the outer layer 4 cm away from the inner layer. Additionally, there are three three-axis reference magnetometers. With the help of a remotely operated position adjustment mechanism, the sensor array can be positioned to provide a uniform short spacing (mean 8.5 mm) between the sensor array and room temperature surface of the dewar. The sensors are connected to superconducting quantum interference devices (SQUIDs) operating at 4.2 K with median sensitivity levels of 7.5 fT/√Hz for the inner and 4 fT/√Hz for the outer layer sensors. SQUID outputs are digitized by a 24-bit acquisition system. A closed-cycle helium recycler provides maintenance-free continuous operation, eliminating the need for helium, with no interruption needed during MEG measurements. BabyMEG with the recycler has been fully operational from March, 2015. Ongoing spontaneous brain activity can be monitored in real time without interference from external magnetic noise sources including the recycler, using a combination of a lightly shielded two-layer magnetically shielded room, an external active shielding, a signal-space projection method, and a synthetic gradiometer approach. Evoked responses in the cortex can be clearly detected without averaging. These new design features and capabilities represent several advances in MEG, increasing the utility of this technique in basic neuroscience as well as in clinical research and patient studies.


Assuntos
Encéfalo/fisiologia , Magnetoencefalografia , Processamento de Sinais Assistido por Computador , Encéfalo/crescimento & desenvolvimento , Pré-Escolar , Humanos , Magnetoencefalografia/instrumentação , Magnetoencefalografia/métodos , Masculino
4.
Front Neuroanat ; 9: 131, 2015.
Artigo em Inglês | MEDLINE | ID: mdl-26528144

RESUMO

BACKGROUND: An understanding of the maturation of auditory cortex responses in typically developing infants and toddlers is needed to later identify auditory processing abnormalities in infants at risk for neurodevelopmental disorders. The availability of infant and young child magnetoencephalography (MEG) systems may now provide near optimal assessment of left and right hemisphere auditory neuromagnetic responses in young populations. To assess the performance of a novel whole-head infant MEG system, a cross-sectional study examined the maturation of left and right auditory cortex responses in children 6- to 59-months of age. METHODS: Blocks of 1000 Hz (1st and 3rd blocks) and 500 Hz tones (2nd block) were presented while MEG data were recorded using an infant/young child biomagnetometer (Artemis 123). Data were obtained from 29 children (11 males; 6- to 59-months). Latency measures were obtained for the first positive-to-negative evoked response waveform complex in each hemisphere. Latency and age associations as well as frequency and hemisphere latency differences were examined. For the 1000 Hz tone, measures of reliability were computed. RESULTS: For the first response-a response with a "P2m" topography-latencies decreased as a function of age. For the second response-a response with a "N2m" topography-no N2m latency and age relationships were observed. A main effect of tone frequency showed earlier P2m responses for 1st 1000 Hz (150 ms) and 2nd 1000 Hz (148 ms) vs. 500 Hz tones (162 ms). A significant main effect of hemisphere showed earlier N2m responses for 2nd 1000 Hz (226 ms) vs. 1st 1000 Hz (241 ms) vs. 500 Hz tones (265 ms). P2m and N2m interclass correlation coefficient latency findings were as follows: left P2m (0.72, p < 0.001), right P2m (0.84, p < 0.001), left N2m (0.77, p < 0.001), and right N2m (0.77,p < 0.01). CONCLUSIONS: Findings of strong age and latency associations, sensitivity to tone frequency, and good test-retest reliability support the viability of longitudinal infant MEG studies that include younger as well as older participants as well as studies examining auditory processing abnormalities in infants at risk for neurodevelopmental disorders.

5.
Front Hum Neurosci ; 8: 99, 2014.
Artigo em Inglês | MEDLINE | ID: mdl-24624069

RESUMO

BACKGROUND: A major motivation in designing the new infant and child magnetoencephalography (MEG) system described in this manuscript is the premise that electrophysiological signatures (resting activity and evoked responses) may serve as biomarkers of neurodevelopmental disorders, with neuronal abnormalities in conditions such as autism spectrum disorder (ASD) potentially detectable early in development. Whole-head MEG systems are generally optimized/sized for adults. Since magnetic field produced by neuronal currents decreases as a function of distance(2) and infants and young children have smaller head sizes (and thus increased brain-to-sensor distance), whole-head adult MEG systems do not provide optimal signal-to-noise in younger individuals. This spurred development of a whole-head infant and young child MEG system - Artemis 123. METHODS: In addition to describing the design of the Artemis 123, the focus of this manuscript is the use of Artemis 123 to obtain auditory evoked neuromagnetic recordings and resting-state data in young children. Data were collected from a 14-month-old female, an 18-month-old female, and a 48-month-old male. Phantom data are also provided to show localization accuracy. RESULTS: Examination of Artemis 123 auditory data showed generalizability and reproducibility, with auditory responses observed in all participants. The auditory MEG measures were also found to be manipulable, exhibiting sensitivity to tone frequency. Furthermore, there appeared to be a predictable sensitivity of evoked components to development, with latencies decreasing with age. Examination of resting-state data showed characteristic oscillatory activity. Finally, phantom data showed that dipole sources could be localized with an error less than 0.5 cm. CONCLUSIONS: Artemis 123 allows efficient recording of high-quality whole-head MEG in infants four years and younger. Future work will involve examining the feasibility of obtaining somatosensory and visual recordings in similar-age children as well as obtaining recordings from younger infants. Thus, the Artemis 123 offers the promise of detecting earlier diagnostic signatures in such neurodevelopmental disorders.