1.
Am J Obstet Gynecol ; 169(6): 1632-5, 1993 Dec.
Artigo em Inglês | MEDLINE | ID: mdl-8267081

RESUMO

OBJECTIVES: Our purpose was to determine whether capacitive currents induced in operative laparoscopes during endoscopic electrosurgery can cause unintentional injury and to measure these currents in the laboratory. STUDY DESIGN: In five anesthetized live pigs we tested whether capacitive currents generated in operative laparoscopes by unipolar instruments cause serosal injury. These currents were then measured in the clinical engineering laboratory. RESULTS: In the setting examined by us serosal injury by capacitive currents occurred only with high generator output power. In the laboratory the maximum power of these capacitive currents measured 2.5 W. CONCLUSIONS: Capacitive coupling between unipolar instruments and 10 mm operating laparoscopes requires relatively high generator output to cause serosal injury. Lower generator output settings may cause injury with electrosurgical generators capable of higher voltages than the units used in this study (Valleylab SSE2L and Force 2).


Assuntos
Condutividade Elétrica , Traumatismos por Eletricidade/etiologia , Eletrocirurgia/efeitos adversos , Laparoscopia/efeitos adversos , Músculos Abdominais/lesões , Animais , Eletrocirurgia/instrumentação , Segurança de Equipamentos , Intestinos/lesões , Laparoscópios , Suínos
SELEÇÃO DE REFERÊNCIAS