Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros










Base de dados
Intervalo de ano de publicação
1.
Rev Sci Instrum ; 83(11): 115107, 2012 Nov.
Artigo em Inglês | MEDLINE | ID: mdl-23206098

RESUMO

We have implemented a new experimental set-up for precise measurements of current fluctuations in three-terminal devices. The system operates at very low temperatures (30 mK) and is equipped with three superconducting quantum interference devices (SQUIDs) as low noise current amplifiers. A SQUID input coil is connected to each terminal of a sample allowing the acquisition of time-dependent current everywhere in the circuit. From these traces, we can measure the current mean value, the noise, and cross-correlations between different branches of a device. In this paper, we present calibration results of noise and cross-correlations obtained using low impedance macroscopic resistors. From these results, we can extract the noise level of the set-up and show that there are no intrinsic correlations due to the measurement scheme. We also studied noise and correlations as a function of a dc current and estimated the electronic temperature of various macroscopic resistors.