Ann Chim ; 92(9): 897-902, 2002 Sep.
Artigo em Inglês | MEDLINE | ID: mdl-12407912

RESUMO

Carbon dioxide monitoring is significant in the environmental field since this gas plays an important role in the greenhouse effect. In order to determine CO2 concentration and to develop simulation models, it is necessary to carry out measurements which are accurate and comparable in time and space, i.e. SI-traceable. Non-dispersive infrared (NDIR) analysers are employed for CO2 measurements, as they are precise and stable. In order to achieve traceability, such instruments have to be characterized and calibrated. At the Istituto di Metrologia "G. Colonnetti"--CNR, a procedure for calibrating NDIR analysers for CO2 at atmospheric level was developed, which enables to calculate a correction for the analyser output. In addition, a complete uncertainty analysis was carried out and a correct traceability chain was established. The goal of the present work is the study of the stability of a NDIR analyser by repeating calibrations during three years and comparing the correction curves obtained to identify a proper re-calibration interval for such analysers. The investigated instrument has good repeatability and reproducibility, hence satisfactory stability during time, as shown by the short-term and long-term compatibility of calibration curves.


Assuntos
Dióxido de Carbono/análise , Espectrofotometria Infravermelho/instrumentação , Atmosfera/análise , Reprodutibilidade dos Testes