Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 2 de 2
Filtrar
Mais filtros










Base de dados
Intervalo de ano de publicação
2.
J Radiol Prot ; 28(4): 511-23, 2008 Dec.
Artigo em Inglês | MEDLINE | ID: mdl-19029585

RESUMO

Currently in the UK, shielding calculations for computed tomography (CT) systems are based on the BIR-IPEM (British Institute of Radiology and Institute of Physics in Engineering in Medicine) working group publication from 2000. Concerns have been raised internationally regarding the accuracy of the dose plots on which this method depends and the effect that new scanner technologies may have. Additionally, more recent shielding methods have been proposed by the NCRP (National Council on Radiation Protection) from the USA. Thermoluminescent detectors (TLDs) were placed in three CT scanner rooms at different positions for several weeks before being processed. Patient workload and dose data (DLP: the dose length product and mAs: the tube current-time product) were collected for this period. Individual dose data were available for more than 95% of patients scanned and the remainder were estimated. The patient workload data were used to calculate expected scatter radiation for each TLD location by both the NCRP and BIR-IPEM methods. The results were then compared to the measured scattered radiation. Calculated scattered air kerma and the minimum required lead shielding were found to be frequently overestimated compared to the measured air kerma, on average almost five times the measured scattered air kerma.


Assuntos
Desenho Assistido por Computador , Análise de Falha de Equipamento/métodos , Modelos Teóricos , Proteção Radiológica/instrumentação , Tomografia Computadorizada por Raios X/instrumentação , Simulação por Computador , Desenho de Equipamento/métodos , Proteção Radiológica/métodos , Reprodutibilidade dos Testes , Sensibilidade e Especificidade , Reino Unido
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...