Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 2 de 2
Filtrar
Mais filtros










Base de dados
Intervalo de ano de publicação
1.
Cardiovasc Intervent Radiol ; 45(8): 1064-1073, 2022 Aug.
Artigo em Inglês | MEDLINE | ID: mdl-35737099

RESUMO

OBJECTIVE: There is no classification system for describing inferior vena cava (IVC) injuries. The objective of this study was to develop a standardized grading system for venographic appearance of the IVC following percutaneous IVC filter retrieval. METHODS: A classification system for the appearance of the IVC on cavograms following percutaneous IVC filter removal was developed consisting of two grading elements; luminal characteristics and extravasation. Luminal narrowing from 0% up to 50% from any cause is grade 1; narrowing between 50 and 99% is grade 2; occlusion is grade 3; and avulsion is grade 4. Absence of extravasation is grade A, contained extravasation is grade B, and free extravasation is grade C. This system was then applied retrospectively to pre- and post-IVC filter retrieval cavograms performed at a single institution from October 2004 through February 2019. RESULTS: 546 retrieval attempts were identified with 509 (93.2%) filters successfully retrieved. 449 cases (88.2%) had both pre-retrieval and post-retrieval imaging appropriate for application of the proposed classification system. Inter-rater reliability was 0.972 for luminal characteristics, 0.967 for extravasation, and 0.969 overall. Consensus grading demonstrated a distribution of 97.3% grade 1, 1.3% grade 2, 1.3% grade 3, and 0.0% grade 4 for post-retrieval luminal characteristics. For extravasation classification, 96.4% of the cases were classified as grade A, 2.7% grade B, and 0.9% grade C. CONCLUSION: A classification system was developed for describing IVC appearance after IVC filter retrieval, and retrospectively validated using a single center dataset.


Assuntos
Filtros de Veia Cava , Veia Cava Inferior , Remoção de Dispositivo/métodos , Humanos , Reprodutibilidade dos Testes , Estudos Retrospectivos , Veia Cava Inferior/diagnóstico por imagem , Veia Cava Inferior/cirurgia
2.
Cardiovasc Diagn Ther ; 7(Suppl 3): S150-S158, 2017 Dec.
Artigo em Inglês | MEDLINE | ID: mdl-29399518

RESUMO

BACKGROUND: To identify risk factors of stent graft thrombosis after transjugular intrahepatic portosystemic shunt (TIPS) creation. METHODS: Patients who underwent TIPS creation between June 2003 and January 2016 and with follow-up assessing stent graft patency were included (n=174). Baseline comorbidities, liver function, procedural details and follow-up liver function tests were analyzed in association with hazards of thrombosis on follow-up. Competing risk cox regression models were used considering liver transplant after TIPS creation as the competing risk variable. RESULTS: One-, 2- and 5-year primary patency rates were 94.1%, 91.7% and 78.2%, respectively. Patient age [sub-hazard ratio (sHR): 1.13; P=0.001], body mass index (BMI) <30 (sHR: 33.08; P=0.008) and a higher post-TIPS portosystemic pressure gradient (sHR: 1.14; P=0.023) were significantly associated with TIPS thrombosis in multivariate analysis. A higher rate of TIPS thrombosis was observed in those for whom the procedure was clinically unsuccessful (P=0.014). A significant increase in incidence of thrombosis was noted with increasing tertiles of post-TIPS portosystemic gradients (P value for trend=0.017). CONCLUSIONS: Older age, lower BMI and higher post-TIPS portosystemic gradients were associated with higher hazards of shunt thrombosis after TIPS creation using stent grafts. Higher rates of shunt thrombosis were seen in patients for whom TIPS creation was clinically unsuccessful. The association between TIPS thrombosis and higher post-TIPS portosystemic gradients may indicate impaired flow through the shunt, a finding which may be technical or anatomic in nature and should be assessed before procedure completion.