1.
J Radiat Res ; 54 Suppl 1: i147-54, 2013 Jul.
Artigo em Inglês | MEDLINE | ID: mdl-23824120

RESUMO

A conceptual design of a mobile isocenter carbon ion gantry was carried out in the framework of the Particle Training Network for European Radiotherapy (PARTNER) and Union of Light Ion Centres in Europe (ULICE) projects. To validate the magnets used in this gantry, Finite Element Method (FEM) simulations were performed with COMSOL multiphysics; the purpose was to evaluate the magnetic field quality and the influence of additional support structures for correctors, 90° bending dipole and quadrupoles, both in dynamic and static regimes. Due to the low ramp rates, the dynamic effects do not disturb the homogeneity and the magnetic field level. The differences between the stationary field and the corresponding dynamic field after the end of the ramps are in the order of 10(-4); it implies that the magnets can be operated without significant field lag at the nominal ramp rate. However, even in static regime the magnetic length of corrector magnet decreases by 5% when the rotator mechanical structure is considered. The simulations suggest an optimization phase of the correctors in the rotator.


Assuntos
Radioterapia com Íons Pesados/instrumentação , Radioterapia com Íons Pesados/métodos , Magnetismo , Neoplasias/radioterapia , Algoritmos , Carbono/química , Simulação por Computador , Desenho de Equipamento , Análise de Elementos Finitos , Humanos , Imãs , Aceleradores de Partículas , Software
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA