Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 2 de 2
Filtrar
Mais filtros










Base de dados
Intervalo de ano de publicação
1.
Platelets ; 21(3): 221-8, 2010.
Artigo em Inglês | MEDLINE | ID: mdl-20158381

RESUMO

Platelet dysfunction is an important cause of bleeding early after cardiac surgery. Whole-blood multiple electrode aggregometry (MEA), investigating the adhesion and aggregation of activated platelets onto metal electrodes, has shown correlations with platelet concentrates transfusion in this setting. Platelet activity in vivo is dependent on shear stress, an aspect that cannot be investigated with MEA, but with the cone and plate(let) analyzer (CPA) Impact-R that measures the interaction of platelets and von Willebrand factor (vWF) in whole blood under shear. We hypothesized that preoperative CPA may show better correlation with platelet concentrates transfusion post-cardiac surgery than MEA, since it is dependent on both platelet activity and platelet interaction with vWF multimers. Blood was obtained preoperatively from 30 patients undergoing aorto-coronary bypass (ACB) and 20 patients with aortic valve (AV) surgery. MEA was performed in hirudin-anticoagulated blood. The Impact-R analyses were performed in blood anticoagulated with hirudin, heparin or the standard anticoagulant citrate. For the light microscopy images obtained, the parameter surface coverage (SC) was calculated. Preoperative Impact-R results were abnormally decreased in AV patients and significantly lower than in ACB patients. For the Impact-R analysis performed in citrated blood, no correlation with platelet concentrates transfusion was observed. In contrast, MEA was comparable between the groups and correlated significantly with intraoperative platelet concentrates transfusion in both groups (rho between -0.47 and -0.62, p < 0.05). Multiple electrode aggregometry appeared more useful and easier to apply than CPA for preoperatively identifying patients with platelet concentrates transfusion in cardiac surgery.


Assuntos
Plaquetas/fisiologia , Ponte de Artéria Coronária , Agregação Plaquetária , Transfusão de Plaquetas , Sistemas Automatizados de Assistência Junto ao Leito , Adolescente , Adulto , Idoso , Idoso de 80 Anos ou mais , Plaquetas/citologia , Adesão Celular , Eletrodos , Feminino , Humanos , Masculino , Pessoa de Meia-Idade , Testes de Função Plaquetária/instrumentação , Testes de Função Plaquetária/métodos , Transfusão de Plaquetas/efeitos adversos , Adulto Jovem , Fator de von Willebrand/metabolismo
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...