1.
Scanning electron microscopy of an irrigation/aspiration tip following a posterior capsular rupture.
J Am Intraocul Implant Soc
; 11(4): 391-2, 1985 Jul.
Artigo
em Inglês
| MEDLINE
| ID: mdl-4030491
RESUMO
The first use of a new irrigation/aspiration tip resulted in a posterior capsular rupture. Scanning electron microscopy of the tip revealed burrs on the aspiration portion of the tip, as well as poor finishing around the irrigation ports.
Assuntos
Extração de Catarata/instrumentação , Cápsula do Cristalino/lesões , Cristalino/lesões , Extração de Catarata/efeitos adversos , Falha de Equipamento , Humanos , Microscopia Eletrônica de Varredura , Ruptura
2.
J Am Intraocul Implant Soc
; 10(1): 81-2, 1984.
Artigo
em Inglês
| MEDLINE
| ID: mdl-6706826
RESUMO
A lathe-cut intraocular lens was hit with the mode-locked YAG laser. The resulting marks (dinks) were examined with scanning electron microscopy and are described.