Dentomaxillofac Radiol ; 44(9): 20150108, 2015.
Artigo em Inglês | MEDLINE | ID: mdl-26090934

RESUMO

OBJECTIVES: To investigate the effect of tube current-exposure time (mAs) reduction on clinical and technical image quality for different CBCT scanners, and to determine preliminary minimally acceptable values for the mAs and contrast-to-noise ratio (CNR) in CBCT. METHODS: A polymethyl methacrylate (PMMA) phantom and an anthropomorphic skull phantom, containing a human skeleton embedded in polyurethane, were scanned using four CBCT devices, including seven exposure protocols. For all protocols, the mAs was varied within the selectable range. Using the PMMA phantom, the CNRAIR was measured and corrected for voxel size. Eight axial slices and one coronal slice showing various anatomical landmarks were selected for each CBCT scan of the skull phantom. The slices were presented to six dentomaxillofacial radiologists, providing scores for various anatomical and diagnostic parameters. RESULTS: A hyperbolic relationship was seen between CNRAIR and mAs. Similarly, a gradual reduction in clinical image quality was seen at lower mAs values; however, for several protocols, image quality remained acceptable for a moderate or large mAs reduction compared with the standard exposure setting, depending on the clinical application. The relationship between mAs, CNRAIR and observer scores was different for each CBCT device. Minimally acceptable values for mAs were between 9 and 70, depending on the criterion and clinical application. CONCLUSIONS: Although noise increased at a lower mAs, clinical image quality often remained acceptable at exposure levels below the manufacturer's recommended setting, for certain patient groups. Currently, it is not possible to determine minimally acceptable values for image quality that are applicable to multiple CBCT models.


Assuntos
Tomografia Computadorizada de Feixe Cônico/métodos , Intensificação de Imagem Radiográfica/métodos , Radiografia Dentária Digital/métodos , Adulto , Pontos de Referência Anatômicos/diagnóstico por imagem , Artefatos , Tomografia Computadorizada de Feixe Cônico/instrumentação , Humanos , Masculino , Mandíbula/diagnóstico por imagem , Maxila/diagnóstico por imagem , Imagens de Fantasmas , Polimetil Metacrilato/química , Doses de Radiação , Radiografia Dentária Digital/instrumentação , Crânio/diagnóstico por imagem , Fatores de Tempo
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...