Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros










Base de dados
Intervalo de ano de publicação
1.
Appl Radiat Isot ; 134: 112-116, 2018 Apr.
Artigo em Inglês | MEDLINE | ID: mdl-29273465

RESUMO

The realization and dissemination of the 68Ga activity measurement unit in Cuba is presented. Firstly, the implementation of the Triple to Double Coincidence Ratio (TDCR) method is described for the calibration of the activity concentration of a 68Ga solution using a Hidex™ commercial liquid scintillation counter and a FORTRAN code developed for the calculation of the 68Ga counting efficiencies in the given measurement system. The relative expanded uncertainty (k = 2) associated with the 68Ga activity concentration obtained with the TDCR method is equal to 2%. With the aim to validate this measurement uncertainty estimate, the method is also applied to an Amersham standard solution of 22Na - a positron emitter with a similar decay mode to the 68Ga disintegration scheme from the point of view of type of emitted particles detected in the measurement system. The observed difference between the measured 22Na activity concentration by the TDCR method and the corresponding reference value traceable to NIST is equal to 0.16%. Outcomes of transferring the 68Ga activity standard, realized with the TDCR method, to the secondary standard radionuclide calibrator Capintec CRC™ 15R and to three radionuclide calibrators used for 68Ga PET applications in a hospital are also shown.


Assuntos
Radioisótopos de Gálio/análise , Serviço Hospitalar de Medicina Nuclear , Calibragem , Cuba , Radioisótopos de Gálio/normas , Padrões de Referência , Reprodutibilidade dos Testes , Incerteza
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...